McKinley G A, Fagerberg D J, George B A
Antimicrob Agents Chemother. 1981 Dec;20(6):777-82. doi: 10.1128/AAC.20.6.777.
Anderson Technical, Inc. has developed rapid, semiautomated equipment for antimicrobial susceptibility testing. A total of 310 fresh clinical bacterial isolates (226 gram-negative and 75 gram-positive) were tested with the Anderson Technical system and compared with those of Micro-Media Systems, Inc. For the gram-positive organisms, 74.3% of the test pairs had identical minimum inhibitory concentration values, whereas 99.1 and 0.9% of the test pairs had minimal inhibitory concentration values differing by less than or equal to 1 and greater than 1 dilution level, respectively. Identical minimum inhibitory concentration values were obtained for 67.2% of the gram-negative test pairs, whereas 97.6 and 2.4% differed by less than or equal to 1 and greater than 1 dilution level, respectively. For all organisms tested, 98.0% differed by less than or equal to 1 dilution level. The Anderson Technical equipment proved to be a rapid and flexible system for microdilution testing.
安德森技术公司已开发出用于抗菌药敏试验的快速半自动设备。使用该安德森技术系统对总共310株新鲜临床细菌分离株(226株革兰氏阴性菌和75株革兰氏阳性菌)进行了测试,并与微媒体系统公司的测试结果进行了比较。对于革兰氏阳性菌,74.3%的测试对具有相同的最低抑菌浓度值,而99.1%和0.9%的测试对的最低抑菌浓度值相差分别小于或等于1个稀释度和大于1个稀释度。67.2%的革兰氏阴性测试对获得了相同的最低抑菌浓度值,而分别有97.6%和2.4%的测试对相差小于或等于1个稀释度和大于1个稀释度。对于所有测试的菌株,98.0%的相差小于或等于1个稀释度。事实证明,安德森技术设备是一种用于微量稀释测试的快速且灵活的系统。