Jones R N, Gavan T L, Barry A L
J Clin Microbiol. 1980 Apr;11(4):426-9. doi: 10.1128/jcm.11.4.426-429.1980.
The Sensititre broth microdilution antibiotic susceptibility testing system was found to be comparable to a reference broth dilution method. Of 3,194 minimum inhibitory concentration data pairs, 93.9% were within the acceptable +/- 1-log2 dilution range. Only rare (0.6%) very major errors were found between methods.