Bobyn J D, Pilliar R M, Cameron H U, Weatherly G C
Clin Orthop Relat Res. 1980 Jul-Aug(150):263-70.
For a study on the effects of pore size variation on the rate of bone growth into porous-surfaced metallic implants and on the strength of fixation resulting from this ingrowth, 4 distinct pore size ranges were prepared on cobalt-base alloy implants with cobalt-base alloy powder particles of different dimensions. The porous implants were placed into canine femurs for periods of 4, 8, and 12 weeks. Mechanical tests were performed to measure the shear strength of fixation of the implants to cortical bone. For implants with powder-made porous surfaces, a pore size range of approximately 50 to 400 microns provided the optimum or maximum fixation strength (17 MPa) in the shortest time period (8 weeks).
为了研究孔径变化对多孔表面金属植入物的骨生长速率以及由此向内生长所导致的固定强度的影响,使用不同尺寸的钴基合金粉末颗粒在钴基合金植入物上制备了4种不同的孔径范围。将多孔植入物植入犬类股骨中4周、8周和12周。进行力学测试以测量植入物与皮质骨固定的剪切强度。对于具有粉末制成的多孔表面的植入物,孔径范围约为50至400微米时,能在最短时间(8周)内提供最佳或最大固定强度(17兆帕)。