Suppr超能文献

新型生物相容性膜2-甲基丙烯酰氧乙基磷酰胆碱-共-甲基丙烯酸正丁酯包覆的二茂铁介导针型葡萄糖传感器的研制

Development of a ferrocene-mediated needle-type glucose sensor covered with newly designed biocompatible membrane, 2-methacryloyloxyethyl phosphorylcholine-co-n-butyl methacrylate.

作者信息

Nishida K, Sakakida M, Ichinose K, Uemura T, Uehara M, Kajiwara K, Miyata T, Shichiri M, Ishihara K, Nakabayashi N

机构信息

Department of Metabolic Medicine, Kumamoto University School of Medicine, Japan.

出版信息

Med Prog Technol. 1995 May;21(2):91-103.

PMID:7565400
Abstract

To prepare the long-life and stable glucose sensor, we developed the ferrocene-mediated needle-type glucose sensor covered with newly designed biocompatible membrane, 2-methacryloyloxyethyl phosphorylcholine-co-n-butyl methacrylate (MPC-co-BMA) membrane. In this membrane, the hydrophilic phosphorylcholine chains were grafted on the hydrophobic polymer surface. 1. The poly(MPC-co-BMA) membrane inhibited platelet activation and protein adhesion on the surface, showing excellent biocompatibility. These results suggested that the hydrophilic phospholipids chains might have the potential for suppressing activation and adsorption of biochemical molecules. 2. The ferrocene-mediated needle-type glucose sensor covered with poly(MPC-co-BMA) membrane achieved excellent results in vitro. Subcutaneous tissue glucose concentrations were measured in a wide range from 1.7 to more than 16.7 mmol/l. The correlation between subcutaneous tissue (Y) and blood (X) glucose concentrations was Y = 1.04X + 0.12 (r = 0.98). The subcutaneous tissue glucose concentrations could be monitored precisely for 7 days without any in vivo calibrations, and for 14 days by introducing in vivo calibrations. We therefore conclude that this sensor is stable and reliable, as compared to any other glucose sensors we developed.

摘要

为制备长寿命且稳定的葡萄糖传感器,我们开发了一种由二茂铁介导的针型葡萄糖传感器,其覆盖有新设计的生物相容性膜,即聚(甲基丙烯酰氧乙基磷酰胆碱 - 共 - 甲基丙烯酸正丁酯)(MPC - co - BMA)膜。在这种膜中,亲水性的磷酰胆碱链接枝在疏水性聚合物表面。1. 聚(MPC - co - BMA)膜抑制血小板活化以及表面的蛋白质粘附,显示出优异的生物相容性。这些结果表明亲水性磷脂链可能具有抑制生化分子活化和吸附的潜力。2. 覆盖有聚(MPC - co - BMA)膜的二茂铁介导的针型葡萄糖传感器在体外取得了优异的结果。皮下组织葡萄糖浓度在1.7至超过16.7 mmol/l的宽范围内进行了测量。皮下组织(Y)与血液(X)葡萄糖浓度之间的相关性为Y = 1.04X + 0.12(r = 0.98)。皮下组织葡萄糖浓度可以在无需任何体内校准的情况下精确监测7天,通过引入体内校准则可监测14天。因此,我们得出结论,与我们开发的任何其他葡萄糖传感器相比,该传感器稳定且可靠。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验