Capes D F, Dunster K R, Sunderland V B, McMillan D, Colditz P B, McDonald C
School of Pharmacy, Curtin University of Technology, Bentley, Western Australia.
Am J Health Syst Pharm. 1995 Aug 1;52(15):1646-53. doi: 10.1093/ajhp/52.15.1646.
Flow continuity of two brands of syringe pumps and four brands of syringes was studied as a possible cause of hemodynamic fluctuations observed in neonates. Cyclical fluctuations were observed in the blood pressure of 14 neonates receiving dopamine infusions by syringe pump at flow rates from 0.2 to 1 mL/hr. Atom 235 and IVAC 770 pumps and various sizes of Terumo, Becton Dickinson, Omnifix, and IVAC syringes were evaluated. Flow continuity was assessed by using a gravimetric technique. The force needed to initiate and maintain syringe plunger motion was also measured. Noncontinuous flow was encountered most commonly with Terumo syringes, which delivered boluses at regular intervals at flow rates up to 5 mL/hr. The interval was dependent on flow rate and was similar to the time between the blood pressure fluctuations observed clinically. The syringe plunger force exhibited regular fluctuations indicative of the plunger sticking, and simultaneous measurement of flow established a direct temporal relationship with boluses. The other syringes tested did not exhibit such fluctuations. No differences were found between the two syringe pumps. Syringe plunger sticking, resulting in intermittent boluses and potential blood pressure fluctuations, may occur at low flow rates and with certain syringe brands. This appeared to be the cause of hemodynamic fluctuations in neonates receiving dopamine infusions.
研究了两种品牌的注射泵和四种品牌的注射器的流量连续性,将其作为新生儿中观察到的血流动力学波动的一个可能原因。在14名通过注射泵以0.2至1毫升/小时的流速接受多巴胺输注的新生儿的血压中观察到周期性波动。对Atom 235和IVAC 770注射泵以及各种尺寸的泰尔茂、贝克顿·迪金森、全塑和IVAC注射器进行了评估。通过重量法评估流量连续性。还测量了启动和维持注射器柱塞运动所需的力。最常遇到非连续流动的是泰尔茂注射器,它在流速高达5毫升/小时时定期输送推注。间隔时间取决于流速,与临床上观察到的血压波动之间的时间相似。注射器柱塞力表现出表明柱塞粘连的规则波动,同时对流量的测量建立了与推注的直接时间关系。测试的其他注射器没有表现出这种波动。两种注射泵之间没有发现差异。注射器柱塞粘连,导致间歇性推注和潜在的血压波动,可能在低流速和使用某些品牌注射器时发生。这似乎是接受多巴胺输注的新生儿血流动力学波动的原因。