Cornish T J, Cotter R J
Department of Pharmacology and Molecular Sciences, Johns Hopkins University School of Medicine, Baltimore, Maryland 21205.
Anal Chem. 1993 Apr 15;65(8):1043-7. doi: 10.1021/ac00056a017.
A compact, laser desorption tandem time-of-flight mass spectrometer is described. The instrument incorporates two dual-stage reflectron analyzers and a collision region for producing product ions by collision-induced dissociation. Selection of ions of a particular mass is accomplished by deflection of ions from stable trajectory angles entering the second reflectron, while the use of a pulsed valve for introduction of the collision gas obviates the need for differential pumping of the collision region. Initial results are presented, as well as a discussion for optimizing the performance of tandem time-of-flight instruments.
描述了一种紧凑型激光解吸串联飞行时间质谱仪。该仪器包含两个双级反射式分析仪和一个用于通过碰撞诱导解离产生产物离子的碰撞区域。通过使离子从进入第二个反射式分析仪的稳定轨迹角度发生偏转来完成对特定质量离子的选择,同时使用脉冲阀引入碰撞气体避免了对碰撞区域进行差动抽气的需要。给出了初步结果,以及关于优化串联飞行时间仪器性能的讨论。