[Quality assurance of mammography in the province of Trent].

作者信息

Valentini A, Nassivera E, Voltolini A, Volani M, Gottardi S

机构信息

Servizio di Fisica Sanitaria, USL C/5, Trento.

出版信息

Radiol Med. 1993 Sep;86(3):240-6.

PMID:8210532
Abstract

In 1991, a second program for quality control in mammography was carried out in the Physics Department of the Trento Hospital. All the six mammographic centers around Trento entered the study and 9 mammographic units were thus considered. Two main types of measurements were obtained: a) determination of the physical variables affecting radiologic system optimization; b) evaluation of image quality and average breast dose, according to DQM program criteria. The film variables, mean gradients and base plus fog level, the half-value layer (HVL) and the entrance exposure free in air were scored as acceptable in all cases. The average breast dose was calculated on the basis of literature data, given the entrance exposure free in air, the HVL and the focus-skin distance values. All the mammographic units but one were scored as "D" (acceptable) on the average breast dose curve, which is defined by the national average value obtained during phase II of the DQM program. This Quality Control program allowed all mammographic units to be optimized and also emphasized the need for periodical quality controls.

摘要

相似文献

1
[Quality assurance of mammography in the province of Trent].
Radiol Med. 1993 Sep;86(3):240-6.
2
[The Dosage and Quality of Imaging program and quality assurance in mammography].
Radiol Med. 1991 Jan-Feb;81(1-2):69-72.
3
[The dose and quality program in mammography (DQM). Results of the study carried out in 20 Fruili-Venezia Giulia centers].
Radiol Med. 1993 Dec;86(6):893-8.
4
[The procedures and results of a quality control program in mammography carried out on a regional basis].
Radiol Med. 1996 Mar;91(3):187-93.
5
Quality assurance in screening mammography.乳腺钼靶筛查中的质量保证
Health Devices. 1990 May-Jun;19(5-6):152-98.
6
Mammographic equipment, technique, and quality control.
Curr Opin Radiol. 1991 Aug;3(4):571-8.
7
[Physics, dosimetry, and quality control in mammography].
Radiol Med. 1993 Nov;86(5):687-94.
8
[A program devoted to dose and quality in mammography (DQM)].[一项致力于乳腺钼靶摄影剂量与质量(DQM)的计划]
Radiol Med. 1986 Mar;72(3):116-20.
9
Implementation of the European protocol for quality control of the technical aspects of mammography screening in Bulgaria.在保加利亚实施欧洲乳腺X线筛查技术方面质量控制协议。
Radiat Prot Dosimetry. 2005;114(1-3):403-5. doi: 10.1093/rpd/nch569.
10
[Mammography -a guidance level and the present situation of mammographic dose-].
Igaku Butsuri. 2002;22(2):65-73.