Ambulatory blood pressure monitoring.

作者信息

Nicholson W R, Matthews J N, O'Sullivan J J, Wren C

机构信息

Regional Cardiothoracic Centre, Freeman Hospital, Newcastle upon Tyne.

出版信息

Arch Dis Child. 1993 Dec;69(6):681-4. doi: 10.1136/adc.69.6.681.

Abstract

Ambulatory blood pressure monitoring (ABPM) in adults is proving to be useful. The aim of this study was to determine if ABPM is accurate in the lower blood pressure range encountered in children and, equally important, whether it is acceptable to children. Thirty one children, between the ages of 6 and 18 years, were assessed using an ambulatory blood pressure monitor that uses an auscultatory method. Blood pressure was measured in the contralateral arm with a mercury sphygmomanometer and an oscillometric device at the beginning and end of the study for comparison. Over a blood pressure range of 90-130 mm Hg systolic and 40-80 mm Hg diastolic, a close agreement was found with the sphygmomanometer; the limits of agreement (+/- 2 SD) were 11.6 mm Hg for systolic blood pressure and 13.6 mm Hg for diastolic blood pressure. The bias was less than 1.0 mm Hg. The ambulatory device was worn by all patients for at least 16 hours with an average of 52 recordings per patient. The majority found the device comfortable to wear and were not woken from sleep.

摘要

事实证明,成人动态血压监测(ABPM)是有用的。本研究的目的是确定ABPM在儿童较低血压范围内是否准确,同样重要的是,它是否为儿童所接受。使用采用听诊法的动态血压监测仪对31名6至18岁的儿童进行了评估。在研究开始和结束时,用汞柱式血压计和示波装置在对侧手臂测量血压以作比较。在收缩压90 - 130 mmHg和舒张压40 - 80 mmHg的血压范围内,发现与血压计的测量结果高度一致;收缩压的一致性界限(±2 SD)为11.6 mmHg,舒张压为13.6 mmHg。偏差小于1.0 mmHg。所有患者佩戴动态监测设备至少16小时,每位患者平均记录52次。大多数人觉得该设备佩戴舒适,睡眠时未被吵醒。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验