Gall K P, Verhey L J, Wagner M
Department of Radiation Oncology, Massachusetts General Hospital, Boston 02114.
Med Phys. 1993 Jul-Aug;20(4):1153-9. doi: 10.1118/1.596969.
For certain external beam radiotherapy procedures, precise alignment of patients with the treatment beam is essential for good treatment outcome. A method has been developed for quickly achieving precise patient alignment with the aid of stereoscopically located fiducial markers. The alignment algorithm is developed from standard rigid body mechanics using closed form solutions, obviating the need for iterative fitting methods. The technique is implemented with a digitizing tablet and plane film radiographs. The accuracy of alignment with this method in phantom studies is better than 1 mm and 1 deg relative to CT data. The repeatability of positioning is 0.5 mm (standard deviation) and 0.36 deg (standard deviation).
对于某些外照射放射治疗程序,患者与治疗束的精确对准对于良好的治疗效果至关重要。已开发出一种借助立体定位基准标记快速实现患者精确对准的方法。对准算法是根据标准刚体力学使用封闭形式解开发的,无需迭代拟合方法。该技术通过数字化平板和平面胶片射线照片来实现。在体模研究中,使用此方法相对于CT数据的对准精度优于1毫米和1度。定位的重复性为0.5毫米(标准差)和0.36度(标准差)。