[不同气管导管的比较。气管套囊密封、峰值居中及术后咽痛发生率]

[A comparison of different endotracheal tubes. Tracheal cuff seal, peak centering and the incidence of postoperative sore throat].

作者信息

Hähnel J, Treiber H, Konrad F, Eifert B, Hahn R, Maier B, Georgieff M

机构信息

Universitätsklinik für Anästhesiologie, Ulm.

出版信息

Anaesthesist. 1993 Apr;42(4):232-7.

PMID:8488995
Abstract

UNLABELLED

The study objective was to evaluate cuff seal, tube tip position and incidence of postoperative sore throat in three different endotracheal tubes: the reusable Rüsch "red rubber" tube (low-volume, high-pressure cuff) and the disposable Rüsch "Super Safety Clear" and Mallinckrodt "lo-pro" tubes (both high-volume, low-pressure cuffs).

METHODS

In a prospective randomized trial, 150 consecutive patients scheduled for minor orthopaedic surgery under standardized general anaesthesia, including standardized cuff pressure control, were allocated to one of the three types of endotracheal tube. Leakage and intratracheal tube tip position were investigated by means of fiberoptic tracheoscopy via the endotracheal tube, preceded by pharyngeal instillation of indigocarmine. The semi-standardized interviews for the recording of postoperative sore throat were performed once a day under double-blind conditions from the day of operation to postoperative day 3. Statistical independence testing was based on the chi 2-test, significance was assumed at P < 0.05.

MAIN RESULTS

A past-cuff leakage of the pharyngeally instilled dye was found in three patients with "red rubber" tubes. In all of these cases the tracheal cross section deviated markedly from circular. The tips of the "Super Safety Clear" tubes tended to be better centred (P = 0.099). Bad tip position, however, had little influence on postoperative throat complaints (P = 0.394). The differences in incidence of postoperative throat complaints between the "red rubber" (48.9%), "Super Safety Clear" (35.6%) and "lo-pro" tubes (34.1%) did not reach the 5% level (P = 0.288). Such complaints, however, were significantly more frequent in females and in obese patients who exceeded their Broca weights by more than 25%. By contrast, postoperative throat complaints did not vary statistically significantly by age group, tracheal cross-sectional shape, use of atropine or duration of intubation.

CONCLUSION

Despite the economic (usable up to 100 times) and ecological (less PVC waste) advantages of the rubber tube, the disposable tubes with high-volume, low-pressure cuffs tended to be superior regarding cuff seal, tube tip position and incidence of postoperative throat complaints. Nevertheless, the continued use of "red rubber" type reusable tubes may be justified in cases where the tube has to stay in place for only a short time.

摘要

未标注

本研究的目的是评估三种不同气管内导管的套囊密封、导管尖端位置及术后咽痛的发生率:可重复使用的吕施“红色橡胶”导管(小容量、高压套囊)以及一次性使用的吕施“超级安全透明”导管和马利克罗斯特“低压力”导管(均为大容量、低压套囊)。

方法

在一项前瞻性随机试验中,150例计划在标准化全身麻醉下行小型骨科手术(包括标准化套囊压力控制)的连续患者被分配至三种气管内导管中的一种。在经气管内导管进行纤维支气管镜检查前,先向咽部滴入靛胭脂,以研究漏气情况和气管内导管尖端位置。从手术当天至术后第3天,在双盲条件下每天进行一次半标准化访谈,以记录术后咽痛情况。基于卡方检验进行统计独立性检验,P<0.05时认为具有统计学意义。

主要结果

在3例使用“红色橡胶”导管的患者中发现咽部滴入的染料在套囊后方有渗漏。在所有这些病例中,气管横截面明显偏离圆形。“超级安全透明”导管的尖端位置往往更居中(P = 0.099)。然而,尖端位置不佳对术后咽痛的影响较小(P = 0.394)。“红色橡胶”导管(48.9%)、“超级安全透明”导管(35.6%)和“低压力”导管(34.1%)术后咽痛发生率的差异未达到5%的水平(P = 0.288)。然而,女性和体重超过布罗卡体重25%以上的肥胖患者出现此类咽痛的频率明显更高。相比之下,术后咽痛在不同年龄组、气管横截面形状、是否使用阿托品或插管持续时间方面在统计学上无显著差异。

结论

尽管橡胶导管具有经济(可重复使用多达100次)和生态(减少聚氯乙烯废物)优势,但大容量低压套囊的一次性导管在套囊密封、导管尖端位置及术后咽痛发生率方面往往更具优势。然而,在导管只需短时间留置的情况下,继续使用“红色橡胶”型可重复使用导管可能是合理的。