Rohrschneider K, Fendrich T, Becker M, Krastel H, Kruse F E, Völcker H E
Department of Ophthalmology, University of Heidelberg, Germany.
Graefes Arch Clin Exp Ophthalmol. 1995 Dec;233(12):743-9. doi: 10.1007/BF00184084.
The purpose of this study was to develop software that allows the performance of routine static threshold perimetry using the scanning laser ophthalmoscope (SLO) and the comparison of the results with conventional computerized cupola perimetry. The original software does not allow performance of static threshold perimetry within a reasonable examination time.
Static perimetry was performed in random order on 50 healthy eyes using our SLO staircase threshold perimetry technique and the Octopus 500 (program 38). We compared the relative sensitivities for each of 25 corresponding visual field locations.
Mean sensitivity in the SLO perimetry amounted to 32.7 dB (range 25-37 dB) while it was 28.7 dB in the Octopus. For all test locations the SLO showed higher dB values on average. The mean difference between both methods was 3.7 +/- 0.8 dB (range 1.4-5.8 dB) when the test locations at the blind spot were excluded (linear regression between the two methods: r = 0.843, P < 0.0001). The mean time interval between two stimulus presentations was 2.5 s with the SLO perimetry.
With the Heidelberg software, automated static threshold perimetry using the SLO is possible within reasonably short examination times. The mean time interval between two test point presentations is about one tenth of that necessary using the original Rodenstock software. There is a systematic difference between SLO and Octopus fields of about 4 dB which was not very much influenced by the stimulus locations.
本研究的目的是开发一种软件,该软件允许使用扫描激光检眼镜(SLO)进行常规静态阈值视野检查,并将结果与传统的电脑化穹窿视野检查进行比较。原始软件不允许在合理的检查时间内进行静态阈值视野检查。
使用我们的SLO阶梯阈值视野检查技术和Octopus 500(程序38)对50只健康眼睛以随机顺序进行静态视野检查。我们比较了25个相应视野位置中每个位置的相对敏感度。
SLO视野检查中的平均敏感度为32.7 dB(范围25 - 37 dB),而Octopus中的平均敏感度为28.7 dB。对于所有测试位置,SLO平均显示出更高的dB值。当排除盲点处的测试位置时,两种方法之间的平均差异为3.7 +/- 0.8 dB(范围1.4 - 5.8 dB)(两种方法之间的线性回归:r = 0.843,P < 0.0001)。SLO视野检查中两次刺激呈现之间的平均时间间隔为2.5秒。
使用海德堡软件,在合理短的检查时间内使用SLO进行自动静态阈值视野检查是可行的。两个测试点呈现之间的平均时间间隔约为使用原始罗敦司得软件所需时间的十分之一。SLO和Octopus视野之间存在约4 dB的系统差异,该差异受刺激位置的影响不大。