Suppr超能文献

蓝黄视野与扫描共焦激光视盘测量的相关性

Correlation of blue-on-yellow visual fields with scanning confocal laser optic disc measurements.

作者信息

Teesalu P, Vihanninjoki K, Airaksinen P J, Tuulonen A, Läärä E

机构信息

Department of Ophthalmology, University of Oulu, Finland.

出版信息

Invest Ophthalmol Vis Sci. 1997 Nov;38(12):2452-9.

PMID:9375562
Abstract

PURPOSE

Visual field defects and changes in the optic nerve head are signs of glaucoma. It has been shown that blue-on-yellow (B-Y) perimetry can reveal visual field defects earlier and shows them larger than does white-on-white (W-W) perimetry. The Heidelberg retina tomograph (HRT) can produce three-dimensional images of the optic disc. The aim of this study was to find out how B-Y perimetry results correlate with optic disc parameters in comparison with W-W perimetry results.

METHODS

One randomly chosen eye was evaluated in each of 40 normal subjects and 37 patients with ocular hypertension and different stages of glaucoma. B-Y and W-W visual fields (program 30-2) were obtained with a Humphrey perimeter. B-Y perimetry results were adjusted for the patient's age and lens transmission index measured with a lens fluorometer. The B-Y visual field adjusted mean deviation (MD) was calculated as the difference between the measured and expected mean sensitivity values, predicted by the regression model fitted in normal subjects. The HRT with software version 1.11 was used to acquire and evaluate topographic measurements of the optic disc.

RESULTS

The cup shape measure showed strongest correlation with the MD of both the B-Y and W-W visual fields. The multiple correlation coefficients from quadratic regression were 0.65 for both visual fields. Except for peripapillary retinal nerve fiber layer measurements, the statistically significant correlations of the B-Y visual field indexes with other HRT parameters were equal to or better than those of W-W perimetry.

CONCLUSIONS

B-Y perimetry MDs are well correlated with optic nerve head parameters measured with the HRT. In early stages of glaucoma, most HRT variables were better correlated with the B-Y MD than with the W-W MD.

摘要

目的

视野缺损和视神经乳头改变是青光眼的体征。研究表明,蓝黄(B-Y)视野检查比白对白(W-W)视野检查能更早发现视野缺损,且显示的缺损范围更大。海德堡视网膜断层扫描仪(HRT)能生成视盘的三维图像。本研究的目的是找出与W-W视野检查结果相比,B-Y视野检查结果与视盘参数之间的相关性。

方法

对40名正常受试者和37名患有不同阶段青光眼的高眼压症患者,每人随机选取一只眼睛进行评估。使用Humphrey视野计获得B-Y和W-W视野(程序30-2)。根据患者年龄和用晶状体荧光计测量的晶状体透光指数对B-Y视野检查结果进行校正。B-Y视野校正平均偏差(MD)计算为测量的平均敏感度值与正常受试者拟合回归模型预测的预期平均敏感度值之间的差值。使用软件版本1.11的HRT获取并评估视盘的地形图测量结果。

结果

杯状测量与B-Y和W-W视野的MD相关性最强。两个视野二次回归的多重相关系数均为0.65。除了视乳头周围视网膜神经纤维层测量外,B-Y视野指数与其他HRT参数的统计学显著相关性等于或优于W-W视野检查。

结论

B-Y视野MD与用HRT测量的视神经乳头参数密切相关。在青光眼早期,大多数HRT变量与B-Y MD的相关性优于与W-W MD的相关性。