Blue-on-yellow (B/Y) perimetry can reveal visual field defects earlier and larger in extent than white-on-white (W/W) perimetry. The Heidelberg Retina Tomograph (HRT) produces a three-dimensional image of the optic disc. The aim of this study was to compare the strength of the association of the B/Y and W/W visual hemifield mean deviation (HMD) variables with the optic nerve head (ONH) morphological variables of the respective area.
We evaluated one randomly chosen eye of 40 normal subjects and 37 patients with ocular hypertension and different stages of glaucoma. The B/Y and W/W visual fields (program 30-2) were obtained with a Humphrey perimeter. Results of both visual fields were adjusted for the patient's age and lens transmission index measured with a lens fluorometer. HMD was calculated as the difference between the measured and expected hemifield mean sensitivity values, predicted by the regression model fitted in our nonglaucomatous subject data. The HRT with the software version 1.11 was used to acquire and evaluate the topographic measurements of the optic disc.
The B/Y and W/W visual field HMDs showed statistically significant correlation with ONH parameters such as cup shape measure (CSM), rim volume, rim area, mean retinal nerve fiber layer (RNFL) thickness and RNFL cross-sectional area. With forward stepwise logistic regression analysis using B/Y hemifield data 38% of the glaucoma patient's normal W/W hemifields were classified abnormal. With the CSM alone in the model 52% of the cases were classified abnormal.
B/Y visual field hemifield mean deviation values correlate well with ONH parameters examined with the HRT.
与白色-白色(W/W)视野检查相比,蓝-黄(B/Y)视野检查能更早且更广泛地发现视野缺损。海德堡视网膜断层扫描仪(HRT)可生成视盘的三维图像。本研究的目的是比较B/Y和W/W视野半视野平均偏差(HMD)变量与相应区域视神经乳头(ONH)形态学变量之间关联的强度。
我们评估了40名正常受试者和37名患有高眼压症及不同阶段青光眼患者中随机选取的一只眼睛。使用Humphrey视野计获得B/Y和W/W视野(程序30-2)。根据患者年龄和用晶状体荧光计测量的晶状体透光指数对两个视野的结果进行校正。HMD计算为测量的半视野平均敏感度值与通过我们非青光眼受试者数据拟合的回归模型预测的预期半视野平均敏感度值之间的差值。使用软件版本1.11的HRT获取并评估视盘的地形图测量值。
B/Y和W/W视野HMD与ONH参数如杯状形态测量(CSM)、盘沿体积、盘沿面积、平均视网膜神经纤维层(RNFL)厚度和RNFL横截面积在统计学上具有显著相关性。使用B/Y半视野数据进行向前逐步逻辑回归分析时,38%青光眼患者的正常W/W半视野被分类为异常。仅将CSM纳入模型时,52%的病例被分类为异常。
B/Y视野半视野平均偏差值与用HRT检查的ONH参数相关性良好。