The FACSCount flow cytometer absolute CD3+, CD3+ CD4+, and CD3+ CD8+ cell counts measured at a field site hospital laboratory in Thailand were compared to FACScan absolute counts obtained at a nearby research laboratory. Correlation coefficients for 208 samples were > or = 0.95. The FACSCount was accurate, and it was easier and less expensive to operate than the FACScan. Additionally, the FACScan-generated lymphocyte percentage value was accurate for use with the FACScan SimulSET software.
在泰国一家战地医院实验室测量的FACSCount流式细胞仪的绝对CD3 +、CD3 + CD4 +和CD3 + CD8 +细胞计数,与在附近研究实验室获得的FACScan绝对计数进行了比较。208个样本的相关系数≥0.95。FACSCount准确,且操作起来比FACScan更简便、成本更低。此外,FACScan生成的淋巴细胞百分比值与FACScan SimulSET软件配合使用时是准确的。