Strauss K, Hannet I, Engels S, Shiba A, Ward D M, Ullery S, Jinguji M G, Valinsky J, Barnett D, Orfao A, Kestens L
Becton Dickinson Immunocytometry Systems Europe, Erembodegem, Belgium.
Cytometry. 1996 Mar 15;26(1):52-9. doi: 10.1002/(SICI)1097-0320(19960315)26:1<52::AID-CYTO8>3.0.CO;2-I.
Flow cytometers are instruments that can determine multiparameter data simultaneously and have a great potential in providing unique information about cells. The FACSCount System is designed as the first dedicated flow cytometer for the clinical laboratory. Its current configuration provides CD4, CD8, and CD3 absolute counts from 100 microliters of whole blood. Adapting the FACSCount System to the clinical setting are minimal sample handling, lysis free cell preparation, automated fluorescence gating, built-in calibrated reference beads, and appropriate error code reporting. Quality control checks ensure that reported CD4 counts, important for clinical follow-up and patient management, are accurate, precise, and reproducible across instruments over time.
流式细胞仪是能够同时测定多参数数据的仪器,在提供有关细胞的独特信息方面具有巨大潜力。FACSCount系统被设计为临床实验室的首款专用流式细胞仪。其当前配置可从100微升全血中得出CD4、CD8和CD3的绝对计数。使FACSCount系统适用于临床环境的特点包括样本处理极少、无需裂解的细胞制备、自动荧光门控、内置校准参考微珠以及适当的错误代码报告。质量控制检查可确保报告的CD4计数准确、精确且在不同仪器间随时间具有可重复性,这对于临床随访和患者管理非常重要。