Attenuation-corrected rest thallium-201/stress technetium 99m sestamibi myocardial SPECT in normals.

作者信息

Matsunari I, Böning G, Ziegler S I, Kosa I, Nekolla S G, Ficaro E P, Schwaiger M

机构信息

Nuklearmedizinische Klinik und Poliklinik der Technischen Universität München, Klinikum rechts der Isar, Munich, Germany.

出版信息

J Nucl Cardiol. 1998 Jan-Feb;5(1):48-55. doi: 10.1016/s1071-3581(98)80010-x.

Abstract

BACKGROUND

Rest thallium-201/stress technetium 99m sestamibi protocol is widely used in the clinical setting. Although attenuation correction (AC) represents an important recent development in cardiac single photon emission computed tomography (SPECT) imaging, adjacent extracardiac activity can affect the myocardial count density distribution on AC images, particularly with 201Tl. The aims of this study were to compare normal distribution between AC rest 201Tl and stress 99mTc-sestamibi SPECT images as well as to evaluate the effect of extracardiac activity on AC SPECT images with 99mTc and 201Tl.

METHODS AND RESULTS

A phantom measurement and a study of 21 patients with low likelihood of coronary artery disease were performed with a triple-head SPECT system equipped with a americium 241 line source. In the phantom study, the presence of extracardiac activity increased the inferior-to-anterior ratios, particularly with 201Tl (1.01 to 1.32). In the clinical data, reduced count density with 201Tl compared to 99mTc-sestamibi was observed in most of the noninferior segments. On an individual segment basis, 37 (20%) of 189 segments from 11 (52%) of 21 subjects showed reduced count density on the 201Tl image compared to 99mTc-sestamibi by >10% of peak activity.

CONCLUSIONS

There is a significant difference in myocardial count density distribution between 99mTc-sestamibi and 201Tl on AC SPECT images, indicating that a careful image interpretation that considers the different normal count density distribution between the tracers and/or a tracer specific normal database is necessary, especially when defect reversibility is of concern. Further work should aim for the incorporation of scatter correction combined with attenuation correction.

摘要

背景

静息铊-201/负荷锝-99m 甲氧基异丁基异腈显像方案在临床中广泛应用。尽管衰减校正(AC)是心脏单光子发射计算机断层扫描(SPECT)成像领域近期的一项重要进展,但心脏外相邻组织的放射性会影响AC图像上心肌计数密度分布,在使用铊-201时尤其如此。本研究的目的是比较AC静息铊-201和负荷锝-99m甲氧基异丁基异腈SPECT图像的正常分布情况,并评估心脏外组织放射性对99mTc和铊-201的AC SPECT图像的影响。

方法与结果

使用配备镅-241线源的三头SPECT系统进行了模型测量以及对21例冠心病可能性较低的患者进行了研究。在模型研究中,心脏外组织放射性的存在增加了下壁与前壁的比值,使用铊-201时尤为明显(从1.01增至1.32)。在临床数据中,在大多数非下壁节段观察到,与锝-99m甲氧基异丁基异腈相比,铊-201的计数密度降低。在个体节段层面,21例受试者中有11例(52%)的189个节段中的37个(20%)在铊-201图像上显示计数密度相较于锝-99m甲氧基异丁基异腈降低超过峰值活性的10%。

结论

在AC SPECT图像上,锝-99m甲氧基异丁基异腈和铊-201之间心肌计数密度分布存在显著差异,这表明在进行图像解读时,尤其是在关注缺损可逆性时,必须仔细考虑不同示踪剂之间正常计数密度分布的差异和/或特定示踪剂的正常数据库。进一步的工作应致力于将散射校正与衰减校正相结合。