Crossland Michael D, Rubin Gary S
Division of Vision Rehabilitation Research, Institute of Ophthalmology, University College London, United Kingdom.
Optom Vis Sci. 2002 Nov;79(11):735-9. doi: 10.1097/00006324-200211000-00011.
To assess fixation stability in patients, a scanning laser ophthalmoscope (SLO) has typically been required. Disadvantages of this technique include the need for a fixed viewing distance and rigid head support. Some modern infrared eyetrackers do not have these constraints. The purpose of this study was to compare fixation stability measurements made with these instruments.
Normal subjects were asked to fixate a 2.5 degrees high cross while fixation was measured with a SLO and an infrared eyetracker. Bivariate contour ellipse areas were calculated.
There was a linear relationship between the bivariate contour ellipse areas measured using each instrument. Bivariate contour ellipse areas returned from the eyetracker were larger. There was no difference in test-retest variability between the instruments.
The eyetracker indicates fixation to be less stable than the SLO does, perhaps because of eye movements to compensate for small head movements. Our eyetracker can be used to analyze fixation when viewing images at any distance, without the need for head immobilization. The eyetracker and the SLO complement each other in the investigation of visual behavior.
为评估患者的注视稳定性,通常需要使用扫描激光检眼镜(SLO)。该技术的缺点包括需要固定的观察距离和刚性的头部支撑。一些现代红外眼动仪则没有这些限制。本研究的目的是比较使用这些仪器进行的注视稳定性测量。
要求正常受试者注视一个2.5度高的十字,同时用SLO和红外眼动仪测量注视情况。计算双变量轮廓椭圆面积。
使用每种仪器测量的双变量轮廓椭圆面积之间存在线性关系。眼动仪返回的双变量轮廓椭圆面积更大。两种仪器在重测变异性方面没有差异。
眼动仪显示注视稳定性不如SLO,这可能是因为眼睛运动会补偿头部的微小移动。我们的眼动仪可用于分析在任何距离观看图像时的注视情况,而无需固定头部。在视觉行为研究中,眼动仪和SLO相互补充。