Sato Shuichi, Kishida Mamoru, Ito Koichi
Department of Periodontology, Nihon University School of Dentistry, Tokyo, Japan.
J Periodontol. 2004 Sep;75(9):1269-73. doi: 10.1902/jop.2004.75.9.1269.
Professional maintenance is as important for patients with dental implants as it is for patients with natural teeth. However, no proper maintenance instruments have been available for implant patients. The purpose of this in vitro study was to compare the effects of a new ultrasonic scaler (VR), a conventional ultrasonic scaler (SP), and a plastic scaler (PS) on titanium surfaces.
To simulate subgingival conditions, the implant healing abutments were connected to acrylic resin blocks with artificial gingiva using silicon impression material. The abutments were painted with ink as an artificial form of debris. The ink was removed with the VR, SP, or PS scaler for 60 seconds under standardized conditions, and the removal rate was calculated. The roughness of the abutment surface was measured with a profilometer and observed by scanning electron microscopy (SEM).
The removal rate using the VR and SP scalers was higher than that using the PS scaler. No significant differences in the surface roughness or SEM observations were found among the VR, SP, or PS scalers.
In this preliminary study, the new ultrasonic scaler and conventional ultrasonic scaler were shown to be useful for removing artificial debris and produced no significant damage to titanium surfaces compared to plastic scalers. We concluded that new and conventional ultrasonic scalers with a non-metal tip would be suitable for implant maintenance.
专业维护对于种植牙患者和天然牙患者同样重要。然而,目前尚无适用于种植牙患者的合适维护器械。本体外研究的目的是比较新型超声洁牙器(VR)、传统超声洁牙器(SP)和塑料洁牙器(PS)对钛表面的影响。
为模拟龈下情况,使用硅橡胶印模材料将种植愈合基台与带有人工牙龈的丙烯酸树脂块相连。用墨水作为人工形式的碎屑涂覆基台。在标准化条件下,使用VR、SP或PS洁牙器去除墨水60秒,并计算去除率。用轮廓仪测量基台表面粗糙度,并用扫描电子显微镜(SEM)观察。
VR和SP洁牙器的去除率高于PS洁牙器。在VR、SP或PS洁牙器之间,表面粗糙度或SEM观察结果无显著差异。
在这项初步研究中,新型超声洁牙器和传统超声洁牙器被证明可有效去除人工碎屑,与塑料洁牙器相比,对钛表面无显著损伤。我们得出结论,带有非金属尖端的新型和传统超声洁牙器适用于种植体维护。