Suppr超能文献

A flexible perforated microelectrode array for extended neural recordings.

作者信息

Boppart S A, Wheeler B C, Wallace C S

机构信息

Department of Electrical and Computer Engineering, University of Illinois, Urbana-Champaign 61801.

出版信息

IEEE Trans Biomed Eng. 1992 Jan;39(1):37-42. doi: 10.1109/10.108125.

Abstract

A flexible and perforated 32-element planar microelectrode array has been fabricated and used to measure evoked potentials in brain slices. Electrodes are spaced 200 microns apart in a 4 x 8 array and are sandwiched between layers of insulating polyimide. The polyimide sandwich is lifted off its substrate, making it flexible so that it could shape to contoured tissues. Prior to lift off, holes are etched to expose recording sites 15 microns in diameter and to create perforations which allow increased circulation of artificial cerebrospinal fluid to the recording surface of the tissue and, hence, increased viability. Comparisons of evoked potentials measured over time showed an average increase of 10 h to the viability of the slice while using the perforated versus nonperforated arrays.