Suppr超能文献

影像科室的防护围裙:制造商声明的铅当量值需验证。

Protective aprons in imaging departments: manufacturer stated lead equivalence values require validation.

作者信息

Finnerty M, Brennan P C

机构信息

UCD School of Diagnostic Imaging, St Anthony's, Dublin, Ireland.

出版信息

Eur Radiol. 2005 Jul;15(7):1477-84. doi: 10.1007/s00330-004-2571-2. Epub 2005 Mar 24.

Abstract

The composition of protective aprons worn by X-ray personnel to shield against secondary radiation is changing. Lead is being replaced by either lead-free or composite (lead with other high atomic numbered elements) materials. These newer aprons are categorised by manufacturers in terms of lead equivalent values, but it is unclear how these stated values compare with actual lead equivalent values. In this work, the actual lead equivalence of 41 protective aprons from four manufacturers, all specified as having 0.25 mm lead equivalence, were investigated with transmission experiments at 70 and 100 kVp. All aprons were in current use. The aprons were screened for defects, and age, weight and design was recorded along with details of associated quality assurance (QA). Out of the 41 protective aprons examined for actual lead equivalence, 73% were outside tolerance levels, with actual levels in some aprons demonstrating less than half of the nominal values. The lack of compatibility between actual and nominal lead equivalent values was demonstrated by aprons from three of the four manufacturers investigated. The area of the defects found on screening of the protective aprons were within recommendations. The results highlight the need for acceptancy and ongoing checks of protective aprons to ensure that radiation exposure of imaging personnel is kept to a minimum.

摘要

X射线工作人员用于防护二次辐射的防护围裙的成分正在发生变化。铅正被无铅或复合(铅与其他高原子序数元素)材料所取代。这些新型围裙由制造商根据铅当量值进行分类,但尚不清楚这些规定值与实际铅当量值相比如何。在这项工作中,通过在70和100 kVp下进行的透射实验,对来自四个制造商的41件防护围裙的实际铅当量进行了研究,所有这些围裙都规定具有0.25 mm铅当量。所有围裙均为当前正在使用的。对围裙进行了缺陷筛查,并记录了其使用年限、重量和设计以及相关质量保证(QA)的详细信息。在检查实际铅当量的41件防护围裙中,73%超出了公差水平,一些围裙的实际水平不到标称值的一半。在所调查的四个制造商中的三个制造商的围裙中,都证明了实际铅当量值与标称值之间缺乏一致性。在防护围裙筛查中发现的缺陷面积在建议范围内。结果突出表明需要对防护围裙进行验收和持续检查,以确保将成像人员的辐射暴露降至最低。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验