Crouter Scott E, Antczak Amanda, Hudak Jonathan R, DellaValle Diane M, Haas Jere D
Division of Nutritional Sciences, Cornell University, 279 MVR, Ithaca, NY 14853, USA.
Eur J Appl Physiol. 2006 Sep;98(2):139-51. doi: 10.1007/s00421-006-0255-0. Epub 2006 Aug 3.
This study examined the accuracy and reliability of the MedGraphics VO2000 (VO2000) portable metabolic system and the ParvoMedics TrueOne 2400 (TrueOne 2400) metabolic cart against the criterion Douglas bag (DB) method. Ten healthy males (age 20 +/- 1.7 years) had their gas exchange variables measured at rest and during cycling at 50, 100, 150, 200, and 250 W. Each stage was 10-12 min. For half of the stage gas exchange was measured with the DB and TrueOne 2400 simultaneously and for the other half of the stage gas exchange was measured with the VO2000. The testing was performed on two separate days and the order in which the equipment was used in each stage was randomized. Reliability between days for V (E) (CV 7.3-8.8%) was similar among devices, however, for VO2, and VCO2 the VO2000 (CV 14.2-15.8%) was less reliable compared to the DB (CV 5.3-6.0%) and TrueOne 2400 (CV 4.7-5.7%). The TrueOne 2400 was not significantly different from the DB at rest or any work rate for V (E), VO2, or VCO2 (P > or = 0.05). The VO2000 was significantly different from the DB for V (E) at 50-100 W, VO2 at rest and 100-250 W, and VCO2 at rest and 200-250 W (all, P < 0.05). The TrueOne 2400 provides accurate and reliable results for the measurement of gas exchange variables. The VO2000 portable metabolic system was less reliable for measuring VO2 and VCO2 and generally overestimates VO2 at most cycling work rates. Further research is needed to confirm the results found with the VO2000.
本研究针对标准道格拉斯袋(DB)法,检验了MedGraphics VO2000便携式代谢系统和ParvoMedics TrueOne 2400代谢车的准确性和可靠性。10名健康男性(年龄20±1.7岁)在静息状态以及50、100、150、200和250瓦功率的骑行过程中测量其气体交换变量。每个阶段持续10 - 12分钟。在每个阶段的一半时间里,同时使用DB和TrueOne 2400测量气体交换,在另一半时间里,使用VO2000测量气体交换。测试在两个不同的日子进行,每个阶段使用设备的顺序是随机的。不同设备之间,V(E)日间可靠性(CV 7.3 - 8.8%)相似,然而,对于VO2和VCO2,VO2000(CV 14.2 - 15.8%)与DB(CV 5.3 - 6.)和TrueOne 24日可靠性(CV)相比,可靠性较低。对于V(E)、VO2或VCO2,TrueOne 2400在静息状态或任何工作强度下与DB均无显著差异(P≥0.05)。对于V(E),VO2000在50 - 100瓦时与DB有显著差异;对于VO2,在静息状态以及100 - 250瓦时与DB有显著差异;对于VCO2,在静息状态以及200 - 250瓦时与DB有显著差异(均为P<0.05)。TrueOne 2400在测量气体交换变量时能提供准确可靠的结果。VO2000便携式代谢系统在测量VO2和VCO2时可靠性较低,并且在大多数骑行工作强度下通常高估VO2。需要进一步研究以证实VO2000的研究结果。