Lavanya Raghavan, Teo Livia, Friedman David S, Aung Han T, Baskaran Mani, Gao Hong, Alfred Tamuno, Seah Steve K, Kashiwagi Kenji, Foster Paul J, Aung Tin
Glaucoma Department, Singapore National Eye Centre, 11 Third Hospital Avenue, Singapore 168751.
Br J Ophthalmol. 2007 Aug;91(8):1023-6. doi: 10.1136/bjo.2006.113761. Epub 2007 Feb 27.
To compare anterior chamber depth measurements by three non-contact devices-the IOLMaster, scanning peripheral anterior chamber depth analyser (SPAC), and Visante anterior segment optical coherence tomography (AS-OCT) METHODS: Prospective, cross sectional study of 497 phakic subjects over 50 years of age attending a community clinic in Singapore. Anterior chamber depth of the right eye was measured using all three techniques by the same investigator. Depth measurements were made from the corneal epithelium to the anterior lens surface. The values obtained were compared using Bland-Altman analysis.
232 men and 265 women were examined (mean (SD) age, 63.4 (7.9) years). Mean anterior chamber depth was 3.08 (0.36) mm with IOLMaster, 3.10 (0.44) mm with SPAC, and 3.14 (0.34) mm with AS-OCT. A significant difference was present between the anterior chamber depth measurements recorded by the three devices (p<0.0001). Mean differences between the measurements were: AS-OCT v IOLMaster, 0.062 (0.007) mm (95% limits of agreement, -0.37 to 0.25 mm) (p<0.0001); AS-OCT v SPAC, 0.035 (0.011) mm (-0.44 to 0.51 mm) (p = 0.0001); SPAC v IOLMaster, 0.027 (0.012) mm (-0.57 to 0.50 mm) (p = 0.027).
AS-OCT gave systematically deeper anterior chamber measurements than SPAC and IOL-Master. However, as the differences found were small they are unlikely to be clinically important.
比较三种非接触式设备(IOLMaster、扫描周边前房深度分析仪(SPAC)和Visante眼前节光学相干断层扫描(AS-OCT))测量前房深度的情况。方法:对新加坡一家社区诊所的497名年龄超过50岁的有晶状体受试者进行前瞻性横断面研究。由同一名研究者使用这三种技术测量右眼的前房深度。深度测量是从角膜上皮到晶状体前表面。使用Bland-Altman分析比较获得的值。
检查了232名男性和265名女性(平均(标准差)年龄,63.4(7.9)岁)。IOLMaster测量的平均前房深度为3.08(0.36)mm,SPAC为3.10(0.44)mm,AS-OCT为3.14(0.34)mm。三种设备记录的前房深度测量值之间存在显著差异(p<0.0001)。测量值之间的平均差异为:AS-OCT与IOLMaster相比,0.062(0.007)mm(95%一致性界限,-0.37至0.25 mm)(p<0.0001);AS-OCT与SPAC相比,0.035(0.011)mm(-0.44至0.51 mm)(p = 0.0001);SPAC与IOLMaster相比,0.027(0.012)mm(-0.57至0.50 mm)(p = 0.027)。
AS-OCT系统测量的前房深度比SPAC和IOLMaster更深。然而,由于发现的差异较小,它们在临床上不太可能具有重要意义。