使用两种光学设备测量的眼前节参数与超声数据的比较。

Anterior segment parameters measured with 2 optical devices compared to ultrasonic data.

作者信息

Nemeth Gabor, Hassan Ziad, Szalai Eszter, Berta Andras, Modis Laszlo

机构信息

Department of Ophthalmology, Medical and Health Science Center, University of Debrecen, Debrecen, Hungary.

出版信息

Eur J Ophthalmol. 2013 Mar-Apr;23(2):177-82. doi: 10.5301/ejo.5000214.

Abstract

PURPOSE

Ophthalmic parameters can be measured with different techniques and devices. Our aim was to measure anterior segment parameters using 2 optical devices and to compare the data to ultrasonic technique.

METHODS

Axial length (AL) was measured with IOLMaster; anterior chamber depth (ACD) was determined using IOLMaster, Visante® optical coherence tomography (OCT), and Pentacam® HR; and central corneal thickness (CCT) were measured with Pentacam® HR and with ultrasound.

RESULTS

In 45 eyes of 45 patients (age 62.73±15.31), the AL was 23.35±1.49 mm. The CCT was 541.7±39.9 µm with Visante OCT, 549.7±39.2 µm with Pentacam® HR, and 534.3±41.3 µm with ultrasonic device. The ACD was 3.0±0.42 mm with IOLMaster, 3.26±0.95 mm with Visante OCT, and 3.06±0.41 mm with Pentacam® HR. In case of CCT measurements, data obtained with Pentacam HR correlated with the results of ultrasonic measurements (r=0.89; p<0.001) and Visante OCT measurements (r=0.88; p<0.001). Regarding ACD results, a significant correlation was observed between Pentacam HR and IOLMaster (r=0.95; p<0.0001) but they did not significantly correlate to Visante OCT data (r=0.23; p=0.12). Correlation was significant between AL and ACD measured with IOLMaster (r=0.67; p<0.001).

CONCLUSIONS

The CCT and ACD data show strong correlation between different devices.

摘要

目的

眼科参数可通过不同技术和设备进行测量。我们的目的是使用两种光学设备测量眼前节参数,并将数据与超声技术进行比较。

方法

使用IOLMaster测量眼轴长度(AL);使用IOLMaster、Visante®光学相干断层扫描(OCT)和Pentacam® HR测定前房深度(ACD);使用Pentacam® HR和超声测量中央角膜厚度(CCT)。

结果

在45例患者(年龄62.73±15.31岁)的45只眼中,AL为23.35±1.49 mm。使用Visante OCT测量的CCT为541.7±39.9 µm,使用Pentacam® HR测量的为549.7±39.2 µm,使用超声设备测量的为534.3±41.3 µm。使用IOLMaster测量的ACD为3.0±0.42 mm,使用Visante OCT测量的为3.26±0.95 mm,使用Pentacam® HR测量的为3.06±0.41 mm。在CCT测量中,Pentacam HR获得的数据与超声测量结果(r=0.89;p<0.001)和Visante OCT测量结果(r=0.88;p<0.001)相关。关于ACD结果,Pentacam HR与IOLMaster之间观察到显著相关性(r=0.95;p<0.0001),但它们与Visante OCT数据无显著相关性(r=0.23;p=0.12)。IOLMaster测量的AL与ACD之间相关性显著(r=0.67;p<0.001)。

结论

CCT和ACD数据显示不同设备之间有很强的相关性。