基于光学相干断层扫描技术的儿童晶状体厚度测量的验证

Validation of optical coherence tomography-based crystalline lens thickness measurements in children.

作者信息

Lehman Bret M, Berntsen David A, Bailey Melissa D, Zadnik Karla

机构信息

The Ohio State University, College of Optometry, Columbus, Ohio 43210-1240, USA.

出版信息

Optom Vis Sci. 2009 Mar;86(3):181-7. doi: 10.1097/OPX.0b013e318198198d.

Abstract

PURPOSE

To evaluate the validity and repeatability of crystalline lens thickness measurements obtained by anterior segment optical coherence tomography (OCT).

METHODS

Forty-seven normal children (mean age, 11.06 +/- 2.30 yr) had their crystalline lens thickness measured with the Visante anterior segment OCT (Carl Zeiss Meditec, Dublin, CA) and with conventional corneal touch A-scan ultransonography (ultrasound) (Humphrey 820). The subjects' right corneas were anesthetized, and their right eyes were cyclopleged. Five ultrasound measurements were recorded per eye, and three Visante OCT measurements were recorded per eye. Thirty-eight subjects had measurements at a second visit where three additional Visante OCT measurements were recorded.

RESULTS

The mean of the differences between the Visante OCT and ultrasound was -0.045 mm (p = 0.017) with 95% limits of agreement from -0.29 to 0.20 mm, indicating that the measurement of crystalline lens thickness was slightly thinner with the Visante OCT. When validity was assessed using only Visante OCT images that contained the corneal reflex, the mean of the differences was 0.019 mm (p = 0.11) with 95% limits of agreement from -0.091 to 0.13 mm. For the repeatability of the Visante OCT, the mean of the differences between visit one and visit two was -0.008 mm (p = 0.25) with 95% limits of agreement from -0.088 to 0.072 mm. Repeatability improved when reassessed using only images that contain the corneal reflex; the mean of the differences was -0.0001 mm (p = 0.97) with 95% limits of agreement from -0.030 to 0.030 mm.

CONCLUSION

The Visante OCT is a non-contact instrument that is simple to use, and it provides valid crystalline lens thickness measurements with excellent repeatability. Validity and repeatability are optimized when the Visante OCT images contain the corneal reflex and a consistent corneal index refraction is applied to the entire image.

摘要

目的

评估眼前节光学相干断层扫描(OCT)测量晶状体厚度的有效性和可重复性。

方法

47名正常儿童(平均年龄11.06±2.30岁)使用Visante眼前节OCT(卡尔蔡司医疗技术公司,加利福尼亚州都柏林)和传统角膜接触式A超超声检查(Humphrey 820)测量晶状体厚度。对受试者的右眼进行麻醉,并使其右眼睫状肌麻痹。每只眼睛记录5次超声测量值,每只眼睛记录3次Visante OCT测量值。38名受试者在第二次就诊时进行测量,额外记录3次Visante OCT测量值。

结果

Visante OCT与超声测量值之间的平均差值为-0.045mm(p = 0.017),95%一致性界限为-0.29至0.20mm,表明使用Visante OCT测量晶状体厚度略薄。仅使用包含角膜反射的Visante OCT图像评估有效性时,平均差值为0.019mm(p = 0.11),95%一致性界限为-0.091至0.13mm。对于Visante OCT的可重复性,第一次就诊和第二次就诊之间的平均差值为-0.008mm(p = 0.25),95%一致性界限为-0.088至0.072mm。仅使用包含角膜反射的图像重新评估时,可重复性有所提高;平均差值为-0.0001mm(p = 0.97),95%一致性界限为-0.030至0.030mm。

结论

Visante OCT是一种非接触式仪器,使用简单,能提供有效的晶状体厚度测量值,且具有出色的可重复性。当Visante OCT图像包含角膜反射并对整个图像应用一致的角膜折射率时,有效性和可重复性可得到优化。

相似文献

引用本文的文献

本文引用的文献

2
Accuracy of cornea and lens biometry using anterior segment optical coherence tomography.
J Biomed Opt. 2007 Nov-Dec;12(6):064023. doi: 10.1117/1.2821844.
9
Repeatability of IOLMaster biometry in children.儿童IOLMaster生物测量的可重复性
Optom Vis Sci. 2004 Nov;81(11):829-34. doi: 10.1097/01.opx.0000145020.33250.c0.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验