MRC Laboratory of Molecular Biology, Hills Road, Cambridge CB2 0QH, UK.
Ultramicroscopy. 2009 Nov;109(12):1411-6. doi: 10.1016/j.ultramic.2009.07.004. Epub 2009 Jul 15.
We compare the direct electron imaging performance at 120keV of a monolithic active pixel sensor (MAPS) operated in a conventional integrating mode with the performance obtained when operated in a single event counting mode. For the combination of sensor and incident electron energy used here, we propose a heuristic approach with which to process the single event images in which each event is renormalised to have an integrated weight of unity. Using this approach we find enhancements in the Nyquist frequency modulation transfer function (MTF) and detective quantum efficiency (DQE) over the corresponding integrating mode values by factors of 8 and 3, respectively.
我们比较了在 120keV 下以传统积分模式工作的单片有源像素传感器 (MAPS) 的直接电子成像性能与以单事件计数模式工作时获得的性能。对于此处使用的传感器和入射电子能量组合,我们提出了一种启发式方法来处理单个事件图像,其中每个事件的重新归一化为具有单位积分权重。使用这种方法,我们发现奈奎斯特调制传递函数 (MTF) 和量子探测效率 (DQE) 的频率调制分别提高了 8 倍和 3 倍。