正常受试者短时长瞬态视觉诱发电位的重复性

Repeatability of short-duration transient visual evoked potentials in normal subjects.

作者信息

Tello Celso, De Moraes Carlos Gustavo V, Prata Tiago S, Derr Peter, Patel Jayson, Siegfried John, Liebmann Jeffrey M, Ritch Robert

机构信息

Department of Ophthalmology, Einhorn Clinical Research Center, New York Eye and Ear Infirmary, 310 East 14th Street, New York, NY 10003, USA.

出版信息

Doc Ophthalmol. 2010 Jun;120(3):219-28. doi: 10.1007/s10633-010-9216-3. Epub 2010 Jan 29.

Abstract

To evaluate the within-session and inter-session repeatability of a new, short-duration transient visual evoked potential (SD-tVEP) device on normal individuals, we tested 30 normal subjects (20/20 visual acuity, normal 24-2 SITA Standard VF) with SD-tVEP. Ten of these subjects had their tests repeated within 1-2 months from the initial visit. Synchronized single-channel EEG was recorded using a modified Diopsys Enfant System (Diopsys, Inc., Pine Brook, New Jersey, USA). A checkerboard stimulus was modulated at two reversals per second. Two different contrasts of checkerboard reversal patterns were used: 85% Michelson contrast with a mean luminance of 66.25 cd/m(2) and 10% Michelson contrast with a mean luminance of 112 cd/m(2). Each test lasted 20 s. Both eyes, independently and together, were tested 10 times (5 times at each contrast level). The following information was identified from the filtered N75-P100-N135 complex: N75 amplitude, N75 latency, P100 amplitude, P100 latency, and Delta Amplitude (N75-P100). The median values for each eye's five SD-tVEP parameters were calculated and grouped into two data sets based on contrast level. Mean age was 27.3 +/- 5.2 years. For OD only, the median (95% confidence intervals) of Delta Amplitude (N75-P100) amplitudes at 10% and 85% contrast were 4.6 uV (4.1-5.9) and 7.1 uV (5.15-9.31). The median P100 latencies were 115.2 ms (112.0-117.7) and 104.0 ms (99.9-106.0). There was little within-session variability for any of these parameters. Intraclass correlation coefficients ranged between 0.64 and 0.98, and within subject coefficients of variation were 3-5% (P100 latency) and 15-30% (Delta Amplitude (N75-P100) amplitude). Bland-Altman plots showed good agreement between the first and fifth test sessions (85% contrast Delta Amplitude (N75-P100) delta amplitude, mean difference, 0.48 mV, 95% CI, -0.18-1.12; 85% contrast P100 latency delay, -0.82 ms, 95% CI, -3.12-1.46; 10% contrast Delta Amplitude (N75-P100) amplitude, 0.58 mV, 95% CI, -0.27-1.45; 10% contrast P100 latency delay, -2.05 mV, 95% CI, -5.12-1.01). The inter-eye correlation and agreement were significant for both SD-tVEP amplitude and P100 latency measurements. For the subset of eyes in which the inter-session repeatability was tested, the intraclass correlation coefficients ranged between 0.71 and 0.86 with good agreement shown on Bland-Altman plots. Short-duration transient VEP technology showed good within-session, inter-session repeatability, and good inter-eye correlation and agreement.

摘要

为评估一种新型短持续时间瞬态视觉诱发电位(SD-tVEP)设备在正常个体中的单次测试内和不同测试间的可重复性,我们对30名正常受试者(视力20/20,24-2 SITA标准视野正常)进行了SD-tVEP测试。其中10名受试者在初次就诊后的1-2个月内重复进行了测试。使用改良的Diopsys Enfant系统(美国新泽西州派恩布鲁克的Diopsys公司)记录同步单通道脑电图。棋盘格刺激以每秒两次反转进行调制。使用了两种不同对比度的棋盘格反转模式:85%的迈克尔逊对比度,平均亮度为66.25 cd/m²,以及10%的迈克尔逊对比度,平均亮度为112 cd/m²。每次测试持续20秒。双眼分别及同时进行了10次测试(每个对比度水平各5次)。从滤波后的N75-P100-N135复合波中识别出以下信息:N75波幅、N75潜伏期、P100波幅、P100潜伏期以及差值波幅(N75-P100)。计算每只眼睛五个SD-tVEP参数的中位数,并根据对比度水平将其分为两个数据集。平均年龄为27.3±5.2岁。仅对于右眼,10%和85%对比度下差值波幅(N75-P100)波幅的中位数(95%置信区间)分别为4.6 μV(4.1-5.9)和7.1 μV(5.15-9.31)。P100潜伏期的中位数分别为115.2毫秒(112.0-117.7)和104.0毫秒(99.9-106.0)。这些参数在单次测试内的变异性很小。组内相关系数在0.64至0.98之间,受试者内变异系数为3%-5%(P100潜伏期)和15%-30%(差值波幅(N75-P100)波幅)。Bland-Altman图显示第一次和第五次测试之间具有良好的一致性(85%对比度差值波幅(N75-P100)差值波幅,平均差异,0.48 mV,95% CI,-0.18-1.12;85%对比度P100潜伏期延迟,-0.82毫秒,95% CI,-3.12-1.46;10%对比度差值波幅(N75-P100)波幅,0.58 mV,95% CI,-0.27-1.45;10%对比度P100潜伏期延迟;-2.05 mV,95% CI,-5.12-1.01)。双眼之间在SD-tVEP波幅和P100潜伏期测量方面的相关性和一致性均显著。对于测试了不同测试间可重复性的眼睛子集,组内相关系数在0.71至0.86之间,Bland-Altman图显示具有良好的一致性。短持续时间瞬态VEP技术显示出良好的单次测试内、不同测试间可重复性,以及良好的双眼间相关性和一致性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/76d0/2869044/17de07be4f7e/10633_2010_9216_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验