To assess and compare the horizontal dimensions of the anterior chamber by different anterior segment imaging techniques.
Eighty eyes of 40 patients without any ocular disease or previous ocular surgery were recruited and three consecutive measurements of internal anterior chamber diameter and white-to-white distance (WTW) parameters were determined.
Mean internal anterior chamber diameter was 11.80 +/- 0.39, 11.56 +/- 0.47, and 11.61 +/- 0.58 mm with Visante-OCT, SL-OCT, and Pentacam, respectively. Mean WTW distance was 11.87 +/- 0.35 and 11.65 +/- 0.32 mm by IOLMaster and OrbscanIIz, respectively. There was no significant difference between Visante-OCT and Pentacam, IOLMaster, and OrbscanIIz (P = .125, .918, and .314). However, detection of mean internal anterior chamber diameter was significantly greater by Visante-OCT than SL-OCT (P = .026). All devices displayed a high intrasession repeatability (repeated measured ANOVA, P > .05).
Internal anterior chamber diameter measurements using Visante-OCT, SL-OCT, and Pentacam, and WTW measurements using IOLMaster and OrbscanIIz were easy to handle and demonstrated good repeatability. Although similar results of horizontal anterior chamber diameter were determined by Visante-OCT, Pentacam, IOLMaster, and OrbscanIIz, the inter-device differences should be considered during clinical practice.
通过不同的眼前节成像技术评估和比较前房的水平维度。
招募40例无任何眼部疾病或既往眼部手术史的患者的80只眼,连续测量前房内径和白对白距离(WTW)参数3次。
使用Visante-OCT、SL-OCT和Pentacam测量的平均前房内径分别为11.80±0.39、11.56±0.47和11.61±0.58mm。使用IOLMaster和OrbscanIIz测量的平均WTW距离分别为11.87±0.35和11.65±0.32mm。Visante-OCT与Pentacam、IOLMaster和OrbscanIIz之间无显著差异(P = 0.125、0.918和0.314)。然而,Visante-OCT测量的平均前房内径显著大于SL-OCT(P = 0.026)。所有设备均显示出较高的测量内重复性(重复测量方差分析,P>0.05)。
使用Visante-OCT、SL-OCT和Pentacam测量前房内径,以及使用IOLMaster和OrbscanIIz测量WTW操作简便,重复性良好。尽管Visante-OCT、Pentacam、IOLMaster和OrbscanIIz测量的前房水平径结果相似,但在临床实践中应考虑不同设备之间的差异。