To assess the reproducibility of the new spectral domain Cirrus high-definition optical coherence tomography (HD-OCT; Carl Zeiss Meditec, Dublin, CA, USA) for analysis of peripapillary retinal nerve fiber layer (RNFL) thickness in healthy eyes.
Thirty healthy Korean volunteers were enrolled. Three optic disc cube 200 x 200 Cirrus HD-OCT scans were taken on the same day in discontinuous sessions by the same operator without using the repeat scan function. The reproducibility of the calculated RNFL thickness and probability code were determined by the intraclass correlation coefficient (ICC), coefficient of variation (CV), test-retest variability, and Fleiss' generalized kappa (kappa).
Thirty-six eyes were analyzed. For average RNFL thickness, the ICC was 0.970, CV was 2.38%, and test-retest variability was 4.5 microm. For all quadrants except the nasal, ICCs were 0.972 or higher and CVs were 4.26% or less. Overall test-retest variability ranged from 5.8 to 8.1 microm. The kappa value of probability codes for average RNFL thickness was 0.690. The kappa values of quadrants and clock-hour sectors were lower in the nasal areas than in other areas.
The reproducibility of Cirrus HD-OCT to analyze peripapillary RNFL thickness in healthy eyes was excellent compared with the previous reports for time domain Stratus OCT. For the calculated RNFL thickness and probability code, variability was relatively higher in the nasal area, and more careful analyses are needed.
评估新型光谱域 Cirrus 高清光学相干断层扫描(HD-OCT;卡尔蔡司医疗科技公司,都柏林,加利福尼亚州,美国)在分析健康眼中的视盘周围视网膜神经纤维层(RNFL)厚度的可重复性。
纳入 30 名健康的韩国志愿者。同一名操作者在无重复扫描功能的情况下,在不同时间段内对同一只眼进行了 3 次视盘立方 200 x 200 Cirrus HD-OCT 扫描。通过组内相关系数(ICC)、变异系数(CV)、测试-重测变异性和 Fleiss 广义kappa(kappa)来确定计算得到的 RNFL 厚度和概率编码的可重复性。
分析了 36 只眼。对于平均 RNFL 厚度,ICC 为 0.970,CV 为 2.38%,测试-重测变异性为 4.5μm。除了鼻侧象限之外,所有象限的 ICC 均在 0.972 或更高,CV 在 4.26%或更低。总体测试-重测变异性范围为 5.8 至 8.1μm。平均 RNFL 厚度概率编码的 kappa 值为 0.690。鼻侧象限的象限和时钟小时区的 kappa 值低于其他区域。
与时间域 Stratus OCT 的先前报告相比,Cirrus HD-OCT 分析健康眼中视盘周围 RNFL 厚度的可重复性非常好。对于计算得到的 RNFL 厚度和概率编码,鼻侧区域的变异性相对较高,需要更仔细的分析。