School of Materials Engineering, Purdue University, West Lafayette, IN, USA.
J Mater Sci Mater Med. 2010 May;21(5):1453-61. doi: 10.1007/s10856-010-4024-6. Epub 2010 Feb 17.
Mechanical properties and microstructure characterization of a series of graded commercial rigid polyurethane foams commonly used to mimic trabecular bone in testing orthopaedic devices is reported. Compressive testing conducted according to ASTM standard F1839-08, which requires large specimens (50.8 mm x 50.8 mm x 25.4 mm blocks) gave elastic modulus and compressive strength values ranging from 115 to 794 MPa and 4.7 to 24.7 MPa, respectively, for foams having densities of 0.240-0.641 g/cm(3). All these results were within the requirements of the specification for the corresponding grades. Compression testing using smaller specimens (7.5 mm diameter x 15 mm) typical of testing bone, gave results in good agreement with those obtained in the standard tests. Microstructural measurements showed the average pore size ranged from 125 to 234 microm for densities ranging from 0.641 to 0.159 g/cm(3), respectively. The relative modulus as a function of relative density of the foams fit well to the model of Gibson and Ashby. Cyclic testing revealed hysteresis in the lower density foams with a loading modulus statistically equivalent to that measured in monotonic testing. Shore DO durometry (hardness) measurements show good correlations to elastic modulus and compressive strength. The results suggest additional parameters to consider for the evaluation of polyurethane foams for bone analog applications.
本文报道了一系列商业硬质聚氨酯泡沫的力学性能和微观结构特征,这些泡沫常用于模拟测试矫形设备中骨小梁的性能。根据 ASTM 标准 F1839-08 进行的压缩测试需要使用大尺寸样品(50.8mmx50.8mmx25.4mm 块状物),得到的弹性模量和压缩强度值范围分别为 115-794MPa 和 4.7-24.7MPa,对应的泡沫密度为 0.240-0.641g/cm³。所有这些结果都符合相应等级规格的要求。使用小尺寸样品(7.5mm 直径 x 15mm)进行的压缩测试,这些样品通常用于测试骨,得到的结果与标准测试中获得的结果非常吻合。微观结构测量显示,平均孔径范围为 125-234μm,对应的密度范围为 0.641-0.159g/cm³。泡沫的相对模量与相对密度的关系很好地符合 Gibson 和 Ashby 的模型。循环测试显示在低密度泡沫中存在滞后现象,其加载模量在统计学上与单调测试中测量的模量相当。邵氏 D 硬度计(硬度)测量结果与弹性模量和压缩强度有很好的相关性。这些结果表明,对于用于骨模拟应用的聚氨酯泡沫的评估,需要考虑其他附加参数。