Liampa Z, Kynigopoulos M, Pallas G, Gerding H
Department of Retinology, Klinik Pallas, Olten, Switzerland.
Klin Monbl Augenheilkd. 2010 Apr;227(4):285-8. doi: 10.1055/s-0029-1245182. Epub 2010 Apr 20.
An estimation of the repeatability of the Lenstar LS 900 biometry parameters and comparison of these measurements to those evaluated with the IOLMaster were carried out.
Axial length, corneal curvature, anterior chamber depth, central corneal thickness and lens thickness measurements were performed in 145 eyes using the Lenstar LS 900, while axial length, corneal curvature and anterior chamber depth were measured using the IOLMaster. The repeatability coefficients and the coefficients of variation were assessed. The Bland-Altman plots and the regression analysis were applied to evaluate the differences between the two methods.
Axial length was the most repeatable parameter and lens thickness the most variable. Axial length and corneal curvature measurements by IOLMaster and Lenstar LS 900 showed a good correlation (mean differences: 0.03 and 0.01 mm). Anterior chamber depth measurements by the two instruments resulted in a difference of 0.2 mm (all p values < 0.0001). Regression analysis showed a highly significant correlation between measurements by the two methods. Correlation coefficients were: axial length 0.999, corneal curvature 0.983, anterior chamber depth 0.925 (p < 0.0001). Bland-Altman plots demonstrated that measurements by the tested devices were closely related without systematic effects in the range of evaluated data.
Lenstar measurements yielded a high repeatability and correlate highly significant with the IOLMaster measurements.
对Lenstar LS 900生物测量参数的重复性进行了评估,并将这些测量结果与使用IOLMaster评估的结果进行了比较。
使用Lenstar LS 900对145只眼睛进行眼轴长度、角膜曲率、前房深度、中央角膜厚度和晶状体厚度测量,同时使用IOLMaster测量眼轴长度、角膜曲率和前房深度。评估了重复性系数和变异系数。应用Bland-Altman图和回归分析来评估两种方法之间的差异。
眼轴长度是最具重复性的参数,而晶状体厚度是最具变异性的参数。IOLMaster和Lenstar LS 900测量的眼轴长度和角膜曲率显示出良好的相关性(平均差异:0.03和0.01毫米)。两种仪器测量的前房深度相差0.2毫米(所有p值<0.0001)。回归分析显示两种方法的测量结果之间具有高度显著的相关性。相关系数分别为:眼轴长度0.999,角膜曲率0.983,前房深度0.925(p<0.0001)。Bland-Altman图表明,在评估数据范围内,测试设备的测量结果密切相关,无系统效应。
Lenstar测量具有高重复性,且与IOLMaster测量结果高度显著相关。