使用非接触式光学低相干反射仪和角膜曲率计进行生物测量和角膜曲率测量的可重复性和可再现性。

Repeatability and reproducibility of biometry and keratometry measurements using a noncontact optical low-coherence reflectometer and keratometer.

机构信息

Department of Ophthalmology, The Keck School of Medicine, University of Southern California, Los Angeles, California, USA.

出版信息

Am J Ophthalmol. 2012 Jan;153(1):55-61.e2. doi: 10.1016/j.ajo.2011.06.012. Epub 2011 Sep 9.

Abstract

PURPOSE

To evaluate the repeatability, reproducibility, or both of the biometry and keratometry measurements obtained by a new optical low-coherence reflectometer and keratometer (Lenstar LS 900, version 1.10; Haag-Streit).

DESIGN

Prospective, comparative, observational study.

SETTING

Private practice, Lynwood, California, and Santa Monica, California.

STUDY POPULATION

The measurements of the second eye to be operated on of 37 patients between October 2010 and January 2011 were analyzed.

OBSERVATION PROCEDURE

The axial length, central corneal thickness, aqueous depth, anterior chamber depth, crystalline lens thickness, white-to-white corneal diameter, as well as the keratometric readings at the flattest meridian, the steepest meridian, the average K, the amount of astigmatism, and the minus astigmatic cylinder axis were measured before surgery on the patient's first eye and were repeated 1 month later before the second eye surgery. main outcome measures: The repeatability and reproducibility of the biometry and keratometry measurements.

RESULTS

Intrasession repeatability and intersession reproducibility were excellent with a very low coefficient of variation and high interclass correlation coefficients for all measured parameters. Bland-Altman plots show good correlation for axial length measurements (95% limits of agreement ranging from -0.056 to +0.04 mm), anterior chamber depth (-0.22 to +0.18 mm), crystalline lens thickness (-0.21 to +0.27 mm), corneal diameter (-0.28 to +0.24 mm), average keratometric readings (-0.56 to +0.47 diopters), and amount of astigmatism (-0.58 to +0.40 diopters).

CONCLUSIONS

The precision of the measurements obtained by the new optical reflectometer and keratometer is extremely high.

摘要

目的

评估新型光学低相干反射仪和角膜曲率计(Lenstar LS 900,版本 1.10;Haag-Streit)测量的生物测量和角膜曲率的重复性或再现性。

设计

前瞻性、对比性、观察性研究。

地点

加利福尼亚州林伍德和圣莫尼卡的私人诊所。

研究人群

2010 年 10 月至 2011 年 1 月期间,对 37 名即将接受第二只眼手术的患者的测量结果进行了分析。

观察程序

在患者第一只眼手术前,测量眼轴长度、中央角膜厚度、房水深度、前房深度、晶状体厚度、角膜白到白直径,以及最平坦子午线、最陡峭子午线、平均 K 值、散光量和负散光柱轴的角膜曲率读数,并在 1 个月后第二只眼手术前重复测量。

主要观察指标

生物测量和角膜曲率测量的重复性和再现性。

结果

所有测量参数的变异系数均很低,组内相关系数均很高,提示各次测量的重复性和各次检查的再现性均非常好。Bland-Altman 图显示眼轴长度(95%一致性界限范围为-0.056 至+0.04mm)、前房深度(-0.22 至+0.18mm)、晶状体厚度(-0.21 至+0.27mm)、角膜直径(-0.28 至+0.24mm)、平均角膜曲率读数(-0.56 至+0.47 屈光度)和散光量(-0.58 至+0.40 屈光度)的相关性良好。

结论

新型光学反射仪和角膜曲率计测量的精度非常高。