Uzunoglu Emel, Eymirli Ayhan, Uyanik Mehmet Özgür, Çalt Semra, Nagas Emre
Department of Endodontics, Faculty of Dentistry, Hacettepe University, Ankara, Turkey.
Restor Dent Endod. 2015 Feb;40(1):44-9. doi: 10.5395/rde.2015.40.1.44. Epub 2014 Nov 5.
This study compared the ability of several techniques to remove calcium hydroxide (CH) from the root canal and determined the influence of CH residues on the accuracy of the electronic apex locator.
Root canals of 90 human maxillary lateral incisors with confirmed true working length (TWL) were prepared and filled with CH. The teeth were randomly assigned to one of the experimental groups according to the CH removal technique (n = 14): 0.9% saline; 0.9% saline + master apical file (MAF); 17% ethylenediamine tetraacetic acid (EDTA); 17% EDTA + MAF; 5.25% sodium hypochlorite (NaOCl); 5.25% NaOCl + MAF. Six teeth were used as negative control. After CH removal, the electronic working length was measured using Root-ZX (Morita Corp.) and compared with TWL to evaluate Root-ZX accuracy. All specimens were sectioned longitudinally, and the area of remaining CH (CH) and total canal area were measured using imaging software.
The EDTA + MAF and NaOCl + MAF groups showed better CH removal than other groups (p < 0.05). Root-ZX reliability to prevent overestimated working length to be > 85% within a tolerance of ± 1.0 mm (p < 0.05). There was strong negative correlation between amount of CH residues and EAL accuracy (r = -0.800 for ± 0.5 mm; r = -0.940 for ± 1.0 mm).
The mechanical instrumentation improves the CH removal of irrigation solutions although none of the techniques removed the dressing completely. Residues of CH medication in root canals affected the accuracy of Root-ZX adversely.
本研究比较了几种技术从根管中去除氢氧化钙(CH)的能力,并确定了CH残留对电子根尖定位仪准确性的影响。
准备90颗已确认真实工作长度(TWL)的人上颌侧切牙根管,并用CH充填。根据CH去除技术将牙齿随机分配到其中一个实验组(n = 14):0.9%生理盐水;0.9%生理盐水+主尖锉(MAF);17%乙二胺四乙酸(EDTA);17% EDTA + MAF;5.25%次氯酸钠(NaOCl);5.25% NaOCl + MAF。6颗牙齿用作阴性对照。去除CH后,使用Root-ZX(森田公司)测量电子工作长度,并与TWL进行比较以评估Root-ZX的准确性。所有标本纵向切片,使用成像软件测量剩余CH的面积(CH)和根管总面积。
EDTA + MAF组和NaOCl + MAF组的CH去除效果优于其他组(p < 0.05)。在±1.0 mm的公差范围内,Root-ZX防止工作长度高估的可靠性> 85%(p < 0.05)。CH残留量与电子根尖定位仪准确性之间存在强负相关(±0.5 mm时r = -0.800;±1.0 mm时r = -0.940)。
尽管没有一种技术能完全去除根管内的封药,但机械预备可提高冲洗液对CH的去除效果。根管内CH药物残留对Root-ZX的准确性有不利影响。