Marler L M, Siders J A, Wolters L C, Pettigrew Y, Skitt B L, Allen S D
Department of Pathology, Indiana University Medical Center, Indianapolis, Indiana 46202-5113.
J Clin Microbiol. 1991 May;29(5):874-8. doi: 10.1128/jcm.29.5.874-878.1991.
The RapID-ANA II System (Innovative Diagnostic Systems, Inc., Atlanta, Ga.) is a recently revised and marketed 4-h system for the identification of anaerobic bacteria. The system was compared with conventional identification methods for its ability to identify 566 clinical anaerobic isolates. Overall, the system identified correctly to genus and species 68% of the total isolates (62% of 204 gram-negative bacilli, 70% of 69 nonsporeforming gram-positive bacilli, 74% of 130 Clostridium isolates, and 72% of 163 anaerobic cocci), without the use of additional tests. With the additional tests suggested by the manufacturer, 78% of the total isolates were identified correctly to species. The routine use of a few simple and practical tests (e.g., egg yolk agar for Clostridium spp.), in addition to the RapID-ANA II, would improve significantly the accuracy of the system in the identification of anaerobic bacteria. This second-generation system offers a number of improvements over the original system, including an updated data base and the option of overnight refrigeration of the system before the addition of reagents.
RapID-ANA II系统(创新诊断系统公司,佐治亚州亚特兰大)是一种最近经过修订并上市的用于鉴定厌氧菌的4小时系统。该系统与传统鉴定方法在鉴定566株临床厌氧菌分离株的能力方面进行了比较。总体而言,该系统在不使用额外检测的情况下,能将68%的分离株正确鉴定到属和种(204株革兰氏阴性杆菌中的62%,69株无芽孢革兰氏阳性杆菌中的70%,130株梭菌分离株中的74%,以及163株厌氧球菌中的72%)。按照制造商建议进行额外检测后,78%的分离株能被正确鉴定到种。除了RapID-ANA II系统外,常规使用一些简单实用的检测方法(如用于梭菌属的蛋黄琼脂),将显著提高该系统鉴定厌氧菌的准确性。这个第二代系统相对于原始系统有许多改进,包括更新的数据库以及在添加试剂前可选择对系统进行过夜冷藏。