Summanen P, Jousimies-Somer H
Anaerobe Reference Unit, National Public Health Institute, Helsinki, Finland.
Eur J Clin Microbiol Infect Dis. 1988 Dec;7(6):771-5. doi: 10.1007/BF01975045.
This study evaluated RapID ANA and API 20 A systems for identification of anaerobic gram-positive and gram-negative bacteria isolated from oral and non-oral infections using standard reference methods. A total of 480 isolates were tested in both systems. The RapID ANA system correctly identified 74% of the strains to species level and 17.5% to genus level; 5% were misidentified. The API 20 A system correctly identified 50% of the strains to species level and 24.5% to genus level; 8% were misidentified and 17.5% could not be identified by the API 20 A system.
本研究使用标准参考方法,评估了RapID ANA和API 20 A系统对从口腔和非口腔感染中分离出的厌氧革兰氏阳性菌和革兰氏阴性菌的鉴定能力。在这两个系统中总共测试了480株分离菌。RapID ANA系统将74%的菌株正确鉴定到种水平,17.5%鉴定到属水平;5%被错误鉴定。API 20 A系统将50%的菌株正确鉴定到种水平,24.5%鉴定到属水平;8%被错误鉴定,17.5%无法被API 20 A系统鉴定。