Cirrus HD-OCT和Topcon SP-3000P测量中央角膜厚度的准确性。

Accuracy of Cirrus HD-OCT and Topcon SP-3000P for measuring central corneal thickness.

作者信息

Calvo-Sanz Jorge A, Ruiz-Alcocer Javier, Sánchez-Tena Miguel A

机构信息

Instituto de Ciencias Visuales, Madrid, Spain.

Universidad Europea, Madrid, Spain.

出版信息

J Optom. 2018 Jul-Sep;11(3):192-197. doi: 10.1016/j.optom.2016.12.004. Epub 2017 Feb 21.

Abstract

PURPOSE

To compare and analyze the interchangeability of three measuring systems, each based on a different technique, for central corneal thickness (CCT) analysis.

METHODS

CCT measurements were measured using optical coherence tomography (OCT), non-contact specular microscopy (NCSM), and ultrasonic pachymetry (USP) in 60 eyes of 60 healthy patients with a mean age of 66.5±15.0 years and a mean spherical equivalent of 0.43±1.14 D. Analysis of variations in measurement concordance and correlation among the three different methods were performed. Comparison of CCT measurements were done using Bland-Altman plots (with bias and 95% confidence intervals), intraclass correlation coefficient (ICC), and paired t-student analysis.

RESULTS

Mean CCT values were: 549.20±26.91μm for USP (range 503-618μm), 514.20±27.49μm for NCSM (range 456-586μm) and 542.80±25.56μm for OCT (range 486-605μm). CCT values obtained with NCMS were significantly lower than those obtained with OCT and USP methods. NCMS CCT value was 36.08±10.72μm lower than USP value (p<0.05), and NCMS CCT value was 7.88±8.86μm lower than OCT value (p<0.05). ICC between USP-NCSM pair was 0.488 and 0.909 between USP-OCT pair.

CONCLUSION

OCT and UPS offered highly comparable results, whereas NCSM offered lower mean CCT values compared to the other two methods. Therefore, NCSM should not be considered a reliable method for measuring CCT and should rather be considered for assessing longitudinal changes in the same patient.

摘要

目的

比较和分析三种基于不同技术的测量系统在中央角膜厚度(CCT)分析中的互换性。

方法

使用光学相干断层扫描(OCT)、非接触式镜面显微镜检查(NCSM)和超声测厚法(USP)对60例平均年龄为66.5±15.0岁、平均等效球镜度为0.43±1.14 D的健康患者的60只眼睛进行CCT测量。对三种不同方法之间测量一致性和相关性的变化进行分析。使用Bland-Altman图(包括偏差和95%置信区间)、组内相关系数(ICC)和配对t检验分析进行CCT测量值的比较。

结果

USP测量的平均CCT值为:549.20±26.91μm(范围503 - 618μm),NCSM为514.20±27.49μm(范围456 - 586μm),OCT为542.80±25.56μm(范围486 - 605μm)。NCSM获得的CCT值显著低于OCT和USP方法获得的值。NCSM的CCT值比USP值低36.08±10.72μm(p<0.05),NCSM的CCT值比OCT值低7.88±8.86μm(p<0.05)。USP - NCSM对之间的ICC为0.488,USP - OCT对之间的ICC为0.909。

结论

OCT和USP提供了高度可比的结果,而与其他两种方法相比,NCSM的平均CCT值较低。因此,NCSM不应被视为测量CCT的可靠方法,而应考虑用于评估同一患者的纵向变化。