Unexpectedly low pulse oximetry measurements associated with variant hemoglobins: a systematic review.

机构信息

Section of Hematology and Oncology, Department of Medicine, Boston University School of Medicine, Boston, MA 02118, USA.

出版信息

Am J Hematol. 2010 Nov;85(11):882-5. doi: 10.1002/ajh.21810.

Abstract

Pulse oximetry estimates arterial blood oxygen saturation based on light absorbance of oxy- and deoxy-hemoglobin at 660 and 940 nm wavelengths. Patients with unexpectedly low SpO₂ often undergo cardio-pulmonary testing to ascertain the cause of their hypoxemia. However, in a subset of patients, a variant hemoglobin is responsible for low SpO₂ measurements. The extent of this problem is unclear. We performed a systematic literature review for reports of low SpO₂ associated with variant hemoglobins. We also reviewed unpublished cases from an academic hemoglobin diagnostic reference laboratory. Twenty-five publications and four unpublished cases were identified, representing 45 patients with low SpO₂ and confirmed variant hemoglobin. Fifty-seven family members of patients had confirmed or suspected variant hemoglobin. Three low oxygen affinity variant hemoglobins had concordantly low SpO₂ and SaO₂. Eleven variant hemoglobins were associated with unexpectedly low SpO₂ measurements but normal SaO₂. Hemoglobin light absorbance testing was reported in three cases, all of which showed abnormal absorption spectra between 600 and 900 nm. Seven other variant hemoglobins had decreased SpO₂, with unreported or uncertain SaO₂. Twenty-one variant hemoglobins were found to be associated with low SpO₂. Most variant hemoglobins were associated with spuriously low SpO₂. Abnormal absorption spectra explain the discrepancy between SpO₂ and SaO(2) for some variants. The differential diagnosis of possible variant hemoglobin ought to be considered in asymptomatic patients found to have unexpectedly low SpO₂. The correct diagnosis will help to spare patients from unnecessary investigations and anxiety.

摘要

脉搏血氧饱和度仪根据 660 和 940nm 波长处氧合和去氧血红蛋白的光吸收率来估计动脉血氧饱和度。SpO₂ 异常低的患者通常需要进行心肺测试以确定其低氧血症的原因。然而,在一部分患者中,异常血红蛋白是导致 SpO₂ 测量值低的原因。这个问题的严重程度尚不清楚。我们对与异常血红蛋白相关的低 SpO₂ 报告进行了系统的文献回顾。我们还审查了学术血红蛋白诊断参考实验室的未发表病例。确定了 25 篇出版物和 4 个未发表病例,涉及 45 例 SpO₂ 低且异常血红蛋白确诊的患者。患者的 57 名家属有确诊或疑似异常血红蛋白。三种低氧亲和力异常血红蛋白的 SpO₂ 和 SaO₂均低。11 种异常血红蛋白与异常低的 SpO₂ 测量值相关,但 SaO₂正常。有 3 例报告了血红蛋白光吸收率测试,所有患者的吸收光谱均在 600nm-900nm 之间出现异常。另外 7 种异常血红蛋白的 SpO₂ 降低,但 SaO₂ 未报告或不确定。发现 21 种异常血红蛋白与低 SpO₂ 相关。大多数异常血红蛋白与 SpO₂ 假性降低相关。异常吸收光谱解释了某些变体的 SpO₂ 和 SaO₂ 之间的差异。对于发现 SpO₂ 异常低但无症状的患者,应考虑可能异常血红蛋白的鉴别诊断。正确的诊断有助于避免患者进行不必要的检查和焦虑。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验