PURPOSE

To investigate the measures of validity for selective or population screening of the time-domain Stratus and the spectral-domain Cirrus Optical Coherence Tomography (OCT) imaging techniques in a population-based sample and in clinical glaucoma patients at different stages of glaucoma.

PATIENTS AND METHODS

A random sample of 307 subjects living in two rural areas in southern Sweden, and a random sample of 394 clinical glaucoma patients were selected. A large battery of examinations, including Stratus and Cirrus OCT was performed. OCT retinal nerve fibre layer (RNFL) thickness analyses for average thickness, quadrant and clock-hour sectors were compared with normative significance limits available in the instruments.

RESULTS

The population-based sample included 129 healthy and nine glaucoma subjects, and the sample of clinical glaucoma patients included 138 patients. Specificity and positive predictive values were generally better with Stratus than for Cirrus, and sensitivity was generally better with Cirrus. With the average RNFL thickness parameter, Stratus reached 100% specificity and a positive predictive value of 100% and 68% sensitivity in the whole group of the clinical glaucoma patients, but sensitivity was only 28% among the earliest stage glaucoma patients. Sensitivity increased considerably when relying on the quadrant sector parameter, while specificity decreased only marginally.

CONCLUSION

Stratus, with high specificity and positive predictive values, seemed to be best of choice for screening purposes, while Cirrus, with high sensitivity, was the better choice for early detection.

摘要

目的

在基于人群的样本和不同阶段青光眼的临床青光眼患者中,研究时域 Stratus 和频域 Cirrus 光学相干断层扫描(OCT)成像技术的选择性或人群筛查的有效性测量方法。

方法

在瑞典南部两个农村地区随机选择了 307 名受试者和 394 名临床青光眼患者的随机样本。进行了大量检查,包括 Stratus 和 Cirrus OCT。将 OCT 视网膜神经纤维层(RNFL)厚度分析的平均值、象限和时钟小时区域与仪器中提供的正常意义限值进行比较。

结果

基于人群的样本包括 129 名健康人和 9 名青光眼患者,临床青光眼患者样本包括 138 名患者。Stratus 的特异性和阳性预测值通常优于 Cirrus,而 Cirrus 的敏感性通常更高。对于平均 RNFL 厚度参数,Stratus 在整个临床青光眼患者组中达到了 100%的特异性和 100%和 68%的阳性预测值,敏感性仅为早期青光眼患者的 28%。当依赖于象限扇区参数时,敏感性显着增加,而特异性仅略有下降。

结论

Stratus 具有高特异性和阳性预测值,似乎是筛查目的的最佳选择,而 Cirrus 具有高灵敏度,是早期检测的更好选择。