用于测量视网膜神经纤维层厚度的频域光学相干断层扫描与时域光学相干断层扫描之间的一致性。

Agreement between spectral-domain and time-domain OCT for measuring RNFL thickness.

作者信息

Vizzeri G, Weinreb R N, Gonzalez-Garcia A O, Bowd C, Medeiros F A, Sample P A, Zangwill L M

机构信息

Hamilton Glaucoma Center, Department of Ophthalmology, University of California San Diego, La Jolla, CA, USA.

出版信息

Br J Ophthalmol. 2009 Jun;93(6):775-81. doi: 10.1136/bjo.2008.150698. Epub 2009 Mar 19.

Abstract

BACKGROUND/AIMS: To evaluate spectral-domain (SD) optical coherence tomography (OCT) reproducibility and assess the agreement between SD-OCT and Time-Domain (TD) OCT retinal nerve fibre layer (RNFL) measurements.

METHODS

Three Cirrus-SD-OCT scans and one Stratus-TD-OCT scan were obtained from Diagnostic Innovations in Glaucoma Study (DIGS) healthy participants and glaucoma patients on the same day. Repeatability was evaluated using Sw (within-subject standard deviation), CV (coefficient of variation) and ICC (intraclass correlation coefficient). Agreement was assessed using correlation and Bland-Altman plots.

RESULTS

16 healthy participants (32 eyes) and 39 patients (78 eyes) were included. SD-OCT reproducibility was excellent in both groups. The CV and ICC for Average RNFL thickness were 1.5% and 0.96, respectively, in healthy eyes and 1.6% and 0.98, respectively, in patient eyes. Correlations between RNFL parameters were strong, particularly for average RNFL thickness (R(2) = 0.92 in patient eyes). Bland-Altman plots showed good agreement between instruments, with better agreement for average RNFL thickness than for sectoral RNFL parameters (for example, at 90 microm average RNFL thickness, 95% limits of agreement were -13.1 to 0.9 for healthy eyes and -16.2 to -0.3 microm for patient eyes).

CONCLUSIONS

SD-OCT measurements were highly repeatable in healthy and patient eyes. Although the agreement between instruments was good, TD-OCT provided thicker RNFL measurements than SD-OCT. Measurements with these instruments should not be considered interchangeable.

摘要

背景/目的:评估频域(SD)光学相干断层扫描(OCT)的可重复性,并评估SD-OCT与时域(TD)OCT测量视网膜神经纤维层(RNFL)之间的一致性。

方法

在同一天从青光眼诊断创新研究(DIGS)的健康参与者和青光眼患者中获取三次Cirrus-SD-OCT扫描和一次Stratus-TD-OCT扫描。使用Sw(受试者内标准差)、CV(变异系数)和ICC(组内相关系数)评估重复性。使用相关性和Bland-Altman图评估一致性。

结果

纳入16名健康参与者(32只眼)和39名患者(78只眼)。两组中SD-OCT的可重复性均极佳。健康眼中平均RNFL厚度的CV和ICC分别为1.5%和0.96,患者眼中分别为1.6%和0.98。RNFL参数之间的相关性很强,尤其是平均RNFL厚度(患者眼中R(2)=0.92)。Bland-Altman图显示仪器之间一致性良好,平均RNFL厚度的一致性优于扇形RNFL参数(例如,平均RNFL厚度为90微米时,健康眼的一致性界限为-13.1至0.9,患者眼为-16.2至-0.3微米)。

结论

SD-OCT测量在健康眼和患者眼中具有高度可重复性。尽管仪器之间的一致性良好,但TD-OCT测量的RNFL厚度比SD-OCT厚。这些仪器的测量结果不应被视为可互换的。

相似文献

引用本文的文献

本文引用的文献