To study the relationship between retinal nerve fiber layer (RNFL) thickness measurements from time domain (Stratus) and spectral domain (Cirrus HD) optical coherence tomography (OCT) in subjects with optic neuritis.
A total of 18 patients who had suffered monocular acute optic neuritis were imaged by a single trained operator using the Stratus-OCT (fast RNFL scan mode) and Cirrus HD-OCT (optic disc cube mode) on the same day. The relationship between RNFL thickness measurements of the two OCTs (overall and by quadrants) were evaluated using intraclass correlation coefficient (ICC) and Bland-Altman plots. Comparisons between eyes having optic neuritis (ON-group) and fellow eyes (fellow-group) were assessed by a generalized estimating equation (GEE) regression model.
For Stratus-OCT, the median average RNFL was 86.3 and 102 μ in the ON-group and in the fellow-group, respectively. For Cirrus-OCT, the corresponding measurements were 81 and 93 μ. Average RNFL thickness as determined by the two OCT devices was well correlated in the ON-group (ρ = 0.906, P< 0.001) and in the fellow-group (ρ = 0.702, P = 0.001). The median signal strength was significantly higher with Cirrus-OCT. Bland-Altman plots showed that Stratus RNFL measurements were larger than Cirrus-OCT, but smaller when average RNFL thickness was very thin (≤ 56 μ).
Although the Cirrus-OCT and Stratus-OCT RNFL thickness measurements are well correlated in patients with optic neuritis, RNFL measurements cannot be directly compared. Clinicians should be aware that measurements are generally higher with Stratus- than with Cirrus-OCT.
研究视神经炎患者时域(Stratus)和频域(Cirrus HD)光学相干断层扫描(OCT)的视网膜神经纤维层(RNFL)厚度测量值之间的关系。
对 18 名单侧急性视神经炎患者进行单眼检查,使用 Stratus-OCT(快速 RNFL 扫描模式)和 Cirrus HD-OCT(视盘立方体模式)由同一位训练有素的操作员进行。使用组内相关系数(ICC)和 Bland-Altman 图评估两种 OCT(整体和象限)的 RNFL 厚度测量值之间的关系。通过广义估计方程(GEE)回归模型评估视神经炎眼(ON 组)和对侧眼(对侧组)之间的比较。
对于 Stratus-OCT,ON 组和对侧组的平均 RNFL 中位数分别为 86.3 和 102μm。对于 Cirrus-OCT,相应的测量值分别为 81 和 93μm。两种 OCT 设备确定的平均 RNFL 厚度在 ON 组(ρ=0.906,P<0.001)和对侧组(ρ=0.702,P=0.001)中相关性良好。Cirrus-OCT 的信号强度明显更高。Bland-Altman 图显示 Stratus RNFL 测量值大于 Cirrus-OCT,但在平均 RNFL 厚度非常薄(≤56μm)时,测量值较小。
尽管视神经炎患者的 Cirrus-OCT 和 Stratus-OCT RNFL 厚度测量值相关性良好,但不能直接比较 RNFL 测量值。临床医生应注意,Stratus-OCT 的测量值通常高于 Cirrus-OCT。