Influenza Other Respir Viruses. 2011 Mar;5(2):104-9. doi: 10.1111/j.1750-2659.2010.00171.x. Epub 2010 Oct 1.
Rapid influenza testing (RFT) allows for a rapid point-of-care diagnosis of influenza. The Quidel QuickVue Influenza A+B test (QuickVue) has a reported manufacturer's sensitivity and specificity of 73% and 96%, respectively, with nasal swabs. However, investigators have shown sensitivities ranging from 22% to 77% in community settings.
The QuickVue rapid influenza test was evaluated in a population of elementary (K-5) school children, using testing in the home, as part of the Pittsburgh Influenza Prevention Project during the 2007-2008 influenza season. The QuickVue test was performed with nasal swab in full accordance with package instructions and compared with the results of nasal swab semi-quantitative RT-PCR.
Sensitivity of the QuickVue was found to be 27% in this sample. There was no statistically valid correlation between the semi-quantitative PCR result and the QuickVue result.
This study is consistent with the low sensitivity of the QuickVue test also reported by others. Viral load, technique, and the use of nasal swabs were examined as contributing factors but were not found to be explanations for this result. Community testing includes patients who are on the lower spectrum of illness which would not be the case in hospital or clinic samples. This suggests that RFT is less sensitive for patients at the lower spectrum of illness, with less severe disease.
快速流感检测(RFT)可实现对流感的即时现场诊断。Quidel QuickVue 流感 A+B 检测(QuickVue)的制造商报告灵敏度和特异性分别为 73%和 96%,检测对象为鼻拭子。然而,研究人员在社区环境中发现其灵敏度范围为 22%至 77%。
在匹兹堡流感预防项目中,于 2007-2008 流感季节,作为家庭检测的一部分,使用鼻拭子对小学(K-5 年级)儿童进行了 QuickVue 快速流感检测。完全按照试剂盒说明书进行鼻拭子 QuickVue 检测,并与鼻拭子半定量 RT-PCR 结果进行比较。
在该样本中,QuickVue 的灵敏度为 27%。QuickVue 结果与半定量 PCR 结果之间无统计学相关性。
本研究结果与其他研究报道的 QuickVue 检测灵敏度较低的结果一致。病毒载量、技术和使用鼻拭子被认为是导致这种结果的因素,但并未对此结果提供解释。社区检测包括处于疾病较轻谱的患者,而医院或诊所样本中则不会出现这种情况。这表明 RFT 对处于疾病较轻谱且病情较轻的患者的敏感性较低。