University Eye Clinic of Trieste, Trieste, Italy.
Graefes Arch Clin Exp Ophthalmol. 2012 Sep;250(9):1395-8. doi: 10.1007/s00417-011-1901-8. Epub 2012 Jan 4.
To investigate optical coherence tomography (OCT) measurements following implantation of the LENTIS Mplus multifocal IOL, compared with a control group.
OCT scans were performed on 50 eyes with the Topcon 3D OCT-1000 in two groups of patients. The first group consisted of patients following implantation of the LENTIS Mplus, and a second group of age-matched control eyes following implantation of a monofocal aspheric IOL. Macular thickness and macular volume values were compared between groups and assessment for any onscreen visible artifacts.
OCT scans were successfully performed in all 50 eyes with no visible artifacts in either group during scan acquisitions. There were no statistically significant differences (p > 0.05) in any measured or calculated macular thickness or volume values between the two groups.
OCT measurements with the Topcon 3D OCT-1000 is possible and free from visible artifacts in eyes which have had the LENTIS Mplus multifocal IOL implanted. Macular thickness and volume values were similar to those of an age-matched control group of monofocal aspheric IOLs.
研究 LENTIS Mplus 多焦点人工晶状体植入术后的光学相干断层扫描(OCT)测量值,并与对照组进行比较。
使用 Topcon 3D OCT-1000 在两组患者的 50 只眼中进行 OCT 扫描。第一组为植入 LENTIS Mplus 的患者,第二组为植入单焦点非球面人工晶状体的年龄匹配对照组。比较两组之间的黄斑厚度和黄斑体积值,并评估任何屏幕上可见的伪影。
在扫描采集过程中,两组均未出现可见伪影,50 只眼的 OCT 扫描均成功完成。两组之间任何测量或计算的黄斑厚度或体积值均无统计学差异(p > 0.05)。
使用 Topcon 3D OCT-1000 进行 OCT 测量,在植入 LENTIS Mplus 多焦点人工晶状体的眼中没有可见伪影。黄斑厚度和体积值与单焦点非球面人工晶状体的年龄匹配对照组相似。