时域和傅里叶域光学相干断层扫描测量角膜生物力学参数图的观察者内重复性和一致性。

Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography.

机构信息

Department of Ophthalmology, University of California, San Francisco, Box 0730, 10 Koret Way K325, San Francisco, CA 94143-0730, USA.

出版信息

Graefes Arch Clin Exp Ophthalmol. 2010 Nov;248(11):1647-56. doi: 10.1007/s00417-010-1360-7. Epub 2010 Mar 30.

Abstract

BACKGROUND

To assess the intra-examiner repeatability and agreement of corneal pachymetry maps measured by time-domain (Visante OCT) and Fourier-domain optical coherence tomography (RTVue OCT).

METHODS

This observational cross-sectional study enrolled 72 eyes from 72 volunteers. Only one eye of each patient was chosen randomly to receive repeated scanning with both devices by the same examiner, in order to test the intra-observer repeatability. The first scan by each device from all enrolled eyes was used to analyze the difference and agreement between the two devices. The agreement between the two devices was analyzed by the Bland-Altman method. Intra-observer repeatability of each OCT device was analyzed by intra-class correlation (ICC).

RESULTS

The mean corneal thickness of the central 2 mm zone was 524.3 ± 35.7 µm and 525.4 ± 35.3 µm by the Visante OCT and the RTVue OCT respectively (p = 0.089). The intra-observer repeatability of the RTVue OCT (ICC = 0.994) was superior to that of the Visante OCT (ICC = 0.989) in the central 2 mm zone. However, the intra-observer repeatability of the RTVue OCT in the pericentral 2 to 5 mm zone was not superior to that of the Visante OCT. Both OCT devices had similar repeatability in the pericentral 2 to 5 mm zone [ICC of the Visante OCT = 0.991; 95% confidence interval (CI) 0.986-0.995), ICC of the RTVue OCT = 0.991; 95% CI 0.985-0.994)).

CONCLUSIONS

The difference in CCT measurement by the Visante OCT and the RTVue OCT is probably too small to influence clinical decision making for refractive surgery and glaucoma management. The RTVue OCT demonstrated better intra-observer repeatability in the central 2 mm zone, which probably was related to its rapid image acquisition capability.

摘要

背景

评估时域(Visante OCT)和傅里叶域光学相干断层扫描(RTVue OCT)测量的角膜生物力学参数图的观察者内重复性和一致性。

方法

本观察性横断面研究纳入了 72 名志愿者的 72 只眼。为了测试观察者内重复性,每只患者的一只眼被随机选择,由同一位检查者使用两种设备进行重复扫描。所有纳入眼的第一扫使用两种设备来分析两种设备之间的差异和一致性。两种设备之间的一致性通过 Bland-Altman 方法进行分析。使用组内相关系数(ICC)分析每种 OCT 设备的观察者内重复性。

结果

中央 2mm 区的平均角膜厚度分别为 524.3μm±35.7μm 和 525.4μm±35.3μm(P=0.089)。RTVue OCT 的观察者内重复性(ICC=0.994)优于 Visante OCT(ICC=0.989)。然而,RTVue OCT 在中央 2 至 5mm 区的观察者内重复性并不优于 Visante OCT。两种 OCT 设备在中央 2 至 5mm 区具有相似的重复性[Visante OCT 的 ICC=0.991;95%置信区间(CI)0.986-0.995),RTVue OCT 的 ICC=0.991;95%CI 0.985-0.994)]。

结论

Visante OCT 和 RTVue OCT 测量 CCT 的差异可能太小,不会影响屈光手术和青光眼管理的临床决策。RTVue OCT 在中央 2mm 区表现出更好的观察者内重复性,这可能与其快速图像采集能力有关。