Hong Jung Taeck, Sung Kyung Rim, Cho Jung Woo, Yun Sung-Cheol, Kang Sung Yong, Kook Michael S
Department of Ophthalmology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
Korean J Ophthalmol. 2012 Feb;26(1):32-8. doi: 10.3341/kjo.2012.26.1.32. Epub 2012 Jan 14.
To evaluate the effect of the scanning laser ophthalmoscope (SLO) guided re-test mode on short- and long-term measurement variability of peripapillary retinal nerve fiber layer (RNFL) thickness obtained by spectral domain-SLO optical coherence tomography (SD-SLO/OCT).
Seventy five healthy eyes were scanned 3 times per day (intra-session variability) by both the SLO guided re-test mode and the independent mode of SD-SLO/OCT. Subjects were scanned 3 times by both modes at visits within a 2-week interval (inter-session variability). For testing longitudinal variability, 3 separate exams were performed over 6 months by both modes. The coefficient of variation (CV), reproducibility coefficient (RC) and intraclass correlation coefficient of RNFL thickness were compared between the two modes.
The intra-session RC and CV ranged from 5.4 to 12.9 microns and 1.76% to 5.72% when measured by independent mode and 5.4 to 12.5 microns and 1.75% to 5.58% by re-test mode, respectively. The inter-session RC and CV ranged from 5.8 to 13.3 microns and 1.89% to 5.78% by independent mode and 5.8 to 12.7 microns and 1.90% to 5.54% by re-test mode, respectively. Intra-session and inter-session variability measurements were not significantly different between the two modes. The longitudinal RC and CV ranged from 8.5 to 19.2 microns and 2.79% to 7.08% by independent mode and 7.5 to 14.4 microns and 2.33% to 6.22% by re-test mode, respectively. Longitudinal measurement variability was significantly lower when measured by the re-test mode compared to the independent mode (average, p = 0.011).
The SLO guided re-test mode for RNFL thickness measurement in SD-SLO/OCT employing a tracking system improved long-term reproducibility by reducing variability induced by inconsistent scan circle placement.
评估扫描激光检眼镜(SLO)引导的重新测试模式对通过光谱域-SLO光学相干断层扫描(SD-SLO/OCT)获得的视乳头周围视网膜神经纤维层(RNFL)厚度的短期和长期测量变异性的影响。
75只健康眼睛通过SLO引导的重新测试模式和SD-SLO/OCT的独立模式每天扫描3次(会话内变异性)。在2周的间隔内,受试者通过两种模式在每次就诊时扫描3次(会话间变异性)。为了测试纵向变异性,两种模式在6个月内进行了3次单独的检查。比较了两种模式下RNFL厚度的变异系数(CV)、再现性系数(RC)和组内相关系数。
通过独立模式测量时,会话内RC和CV分别为5.4至12.9微米和1.76%至5.72%,通过重新测试模式测量时分别为5.4至12.5微米和1.75%至5.58%。通过独立模式测量时,会话间RC和CV分别为5.8至13.3微米和1.89%至5.78%,通过重新测试模式测量时分别为5.8至12.7微米和1.90%至5.54%。两种模式下的会话内和会话间变异性测量没有显著差异。通过独立模式测量时,纵向RC和CV分别为8.5至19.2微米和2.79%至7.08%,通过重新测试模式测量时分别为7.5至14.4微米和2.33%至6.22%。与独立模式相比,通过重新测试模式测量时纵向测量变异性显著更低(平均值,p = 0.011)。
在采用跟踪系统的SD-SLO/OCT中,用于RNFL厚度测量的SLO引导的重新测试模式通过减少由不一致的扫描圈放置引起的变异性提高了长期再现性。