苏丹喀土穆医院孕妇血红蛋白水平的 HemoCue®血红蛋白计和自动化血液分析仪比较。

Comparison of HemoCue® hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan.

机构信息

Faculty of Medicine, University of Khartoum, P,O, Box 102, Khartoum, Sudan.

出版信息

Diagn Pathol. 2012 Mar 21;7:30. doi: 10.1186/1746-1596-7-30.


DOI:10.1186/1746-1596-7-30
PMID:22436620
原文链接:https://pmc.ncbi.nlm.nih.gov/articles/PMC3342090/
Abstract

BACKGROUND: Assessment of hemoglobin is one of the most reliable indicators for anemia, and is widely used to screen for anemia among pregnant women. The HemoCue® has been widely used for as a point-of-care device for hemoglobin estimation in health facilities. Previous studies showed contradictory results regarding the accuracy of HemoCue®. METHODS: This was a hospital-based cross sectional study carried- out among pregnant women at Khartoum hospital in Sudan to find out whether the measurement of hemoglobin concentration by HemoCue® using venous or capillary samples was comparable to that of the automated hematology analyzer as standard. Bland and Altman method was used to compare the measurements with an acceptable difference of ± 1.0 g/dl. RESULTS: Among the 108 subjects in this study the mean (SD) level of hemoglobin level using HemoCue® venous sample, HemoCue® capillary sample and automated hematology analyzer were 12.70 (1.77), 12.87 (2.04) and 11.53 (1.63) g/dl, respectively. Although the correlations between the measurements were all significant there was no agreement between HemoCue® and automated hematology analyzer. The bias + SD (limits of agreement) for HemoCue® venous versus hematology analyzer was 1.17 ± 1.57 (-1.97, 4.31) g/dl, HemoCue® capillary versus hematology analyzer was 1.34 ± 1.85 (-2.36, 5.04) g/dl, and HemoCue® venous versus HemoCue® capillary samples was 017 ± 1.90 and (3.97-3.63) g/dl. CONCLUSION: Hemoglobin concentration assessment by HemoCue® using either venous or capillary blood samples has shown unacceptable agreement with automated hematology analyzer. VIRTUAL SLIDES: The virtual slide(s) for this article can be found here: http://www.diagnosticpathology.diagnomx.eu/vs/8797022296725036.

摘要

背景:血红蛋白评估是贫血最可靠的指标之一,广泛用于孕妇贫血筛查。HemoCue®已广泛用于医疗机构的血红蛋白即时检测。先前的研究表明 HemoCue®的准确性存在矛盾的结果。

方法:这是一项在苏丹喀土穆医院进行的基于医院的横断面研究,旨在确定使用 HemoCue®静脉或毛细血管样本测量血红蛋白浓度是否与自动化血液分析仪作为标准相媲美。采用 Bland 和 Altman 方法比较测量值,可接受的差异为±1.0 g/dl。

结果:在这项研究的 108 名受试者中,HemoCue®静脉样本、HemoCue®毛细血管样本和自动化血液分析仪的平均(SD)血红蛋白水平分别为 12.70(1.77)、12.87(2.04)和 11.53(1.63)g/dl。虽然测量值之间的相关性均具有统计学意义,但 HemoCue®与自动化血液分析仪之间没有一致性。HemoCue®静脉与血液分析仪的偏差+SD(一致性界限)为 1.17±1.57(-1.97,4.31)g/dl,HemoCue®毛细血管与血液分析仪的偏差+SD 为 1.34±1.85(-2.36,5.04)g/dl,HemoCue®静脉与 HemoCue®毛细血管样本的偏差+SD 为 017±1.90(3.97-3.63)g/dl。

结论:使用 HemoCue®静脉或毛细血管血样评估血红蛋白浓度与自动化血液分析仪的一致性不可接受。

虚拟幻灯片:本文的虚拟幻灯片可以在此处找到:http://www.diagnosticpathology.diagnomx.eu/vs/8797022296725036。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4cee/3342090/17fdff5ade80/1746-1596-7-30-3.jpg
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4cee/3342090/7c712921242e/1746-1596-7-30-1.jpg
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4cee/3342090/2a69cc1b1bc4/1746-1596-7-30-2.jpg
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4cee/3342090/17fdff5ade80/1746-1596-7-30-3.jpg
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4cee/3342090/7c712921242e/1746-1596-7-30-1.jpg
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4cee/3342090/2a69cc1b1bc4/1746-1596-7-30-2.jpg
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4cee/3342090/17fdff5ade80/1746-1596-7-30-3.jpg

相似文献

[1]
Comparison of HemoCue® hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan.

Diagn Pathol. 2012-3-21

[2]
Accuracy of noninvasive and invasive point-of-care total blood hemoglobin measurement in an outpatient setting.

Postgrad Med. 2012-7

[3]
Non-invasive hemoglobin measurement devices require refinement to match diagnostic performance with their high level of usability and acceptability.

PLoS One. 2021

[4]
Hemoglobin point-of-care testing in rural Gambia: Comparing accuracy of HemoCue and Aptus with an automated hematology analyzer.

PLoS One. 2020-10-1

[5]
Comparison between the HemoCue and an automated counter for measuring hemoglobin.

Rev Saude Publica. 2004-8

[6]
Assessing Accuracy and Precision of Hemoglobin Determination in Venous, Capillary Pool, and Single-Drop Capillary Blood Specimens Using three Different HemoCue® Hb Models: The Multicountry Hemoglobin Measurement (HEME) Study.

J Nutr. 2024-7

[7]
Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room.

J Clin Monit Comput. 2016-12

[8]
Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer.

Int J Lab Hematol. 2014-2

[9]
Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement.

Anesthesiology. 2011-9

[10]
Hemoglobin measurement in venous blood compared with pooled and single-drop capillary blood: a method-comparison study in a controlled and survey setting in Uganda among children and women.

Am J Clin Nutr. 2024-4

引用本文的文献

[1]
Enhancing anemia diagnostics and accessibility in India: a policy recommendation for effective anemia management.

Front Health Serv. 2025-5-15

[2]
Comparing HemoCue® and Quantitative Buffy Coat® and Coulter Counter-measured haemoglobin concentrations in African children with acute uncomplicated malaria: a Bland-Altman analysis.

Malar J. 2025-3-11

[3]
Evaluating the Feasibility and Cost-Effectiveness of EzeCheck for Hemoglobin Screening Across India.

Cureus. 2024-11-3

[4]
Non-invasive Hemoglobin Screening for Detection of Postpartum Anemia.

Womens Health Rep (New Rochelle). 2024-7-19

[5]
Emerging Evidence and Critical Issues with the Use of Single-Drop Capillary Blood for the Measurement of Hemoglobin Concentration in Population-Level Anemia Surveys.

Adv Nutr. 2024-10

[6]
Performance of the non-invasive point-of-care device, EzeCheck, for haemoglobin assessment in adults and children in community and institutional care settings.

PLOS Digit Health. 2024-5-8

[7]
Improving Anemia Assessment in Clinical and Public Health Settings.

J Nutr. 2023-12

[8]
Diagnosing anemia: Challenges selecting methods, addressing underlying causes, and implementing actions at the public health level.

Ann N Y Acad Sci. 2023-6

[9]
Anemia Among Women Using Family Planning at Public Health Facilities in Ambo Town, Central Ethiopia: Multi-Center Cross-Sectional Study.

J Blood Med. 2023-2-8

[10]
Effect of a fifth round of seasonal malaria chemoprevention in children aged 5-14 years in Dangassa, an area of long transmission in Mali.

Parasite Epidemiol Control. 2023-1-2

本文引用的文献

[1]
Hemoglobin estimation by the HemoCue® portable hemoglobin photometer in a resource poor setting.

BMC Clin Pathol. 2011-4-21

[2]
A perspective of the epidemiology of malaria and anaemia and their impact on maternal and perinatal outcomes in Sudan.

J Infect Dev Ctries. 2011-3-2

[3]
Determination of capillary hemoglobin levels using the HemoCue system in intensive care patients.

J Crit Care. 2010-10-30

[4]
Anaemia and low birth weight in Medani, Hospital Sudan.

BMC Res Notes. 2010-6-28

[5]
Anaemia and stillbirth in Kassala Hospital, Eastern Sudan.

J Trop Pediatr. 2010-4-28

[6]
Evaluation of a portable hemoglobin photometer in pregnant women in a high altitude area: a pilot study.

BMC Public Health. 2009-7-11

[7]
Reliability of HemoCue in patients with gastrointestinal bleeding.

Intensive Care Med. 2007-2

[8]
Hemoglobin screening in prospective blood donors: comparison of different blood samples and different quantitative methods.

Transfus Apher Sci. 2005-8

[9]
Comparison between the HemoCue and an automated counter for measuring hemoglobin.

Rev Saude Publica. 2004-8

[10]
The world health report 2002 - reducing risks, promoting healthy life.

Educ Health (Abingdon). 2003-7

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

推荐工具

医学文档翻译智能文献检索