Hatef Elham, Khwaja Afsheen, Rentiya Zubir, Ibrahim Mohamed, Shulman Matthew, Turkcuoglu Peykan, Sepah Yasir, Wang Jianmin, Channa Roomasa, Bittencourt Millena, Akhtar Abeer, Heo Jangwon, Do Diana V, Nguyen Quan Dong
Retinal Imaging Research and Reading Center (RIRRC), Wilmer Eye Institute, Johns Hopkins Hospital, 600 North Wolfe Street, Maumenee 745, Baltimore, MD 21287, USA.
J Ophthalmol. 2012;2012:354783. doi: 10.1155/2012/354783. Epub 2012 Jul 25.
Purpose. To evaluate macular thickness, agreement, and intraclass repeatability in three optical coherence tomography (OCT) devices: the time domain (TD) Stratus OCT and two spectral domain (SD) OCTs, Spectralis and Cirrus SD-OCT, in eyes with macular edema secondary to diabetic retinopathy (DR) and retinal vein occlusion (VO). Methods. In a prospective observational study at a university-based retina practice, retinal thickness tomography was performed simultaneously for fifty-eight patients (91 eyes) with DR and VO employing a time domain and two spectral domain OCTs. Agreement in macular measurements was assessed by constructing Bland-Altman plots. Intraclass repeatability was assessed by intraclass correlation coefficients (ICCs). Results. Based on the Bland-Altman plots for central macular thickness, there was low agreement between the measurements of Cirrus SD-OCT and Stratus OCT, Spectralis OCT and Stratus OCT, as well as Spectralis OCT and Cirrus SD-OCT among DR and RVO patients. All three devices demonstrated high intraclass repeatability, with ICC of 98% for Stratus OCT, 97% for Cirrus SD-OCT, and 100% for Spectralis OCT among DR patients. The ICC was 97% for Stratus OCT, 79% for Cirrus SD-OCT, and 91% for Spectralis OCT among RVO patients. Conclusion. There are low agreements among interdevice measurements. However, intraclass repeatability is high in both TD and SD-OCT devices.
目的。评估三种光学相干断层扫描(OCT)设备,即时域(TD)Stratus OCT和两种频域(SD)OCT(Spectralis和Cirrus SD - OCT),对糖尿病性视网膜病变(DR)和视网膜静脉阻塞(VO)继发黄斑水肿患者的黄斑厚度、一致性和组内重复性。方法。在一所大学视网膜专科门诊进行的一项前瞻性观察研究中,使用时域和两种频域OCT对58例(91只眼)DR和VO患者同时进行视网膜厚度断层扫描检查。通过构建Bland - Altman图评估黄斑测量的一致性。通过组内相关系数(ICC)评估组内重复性。结果基于中心黄斑厚度的Bland - Altman图,在DR和RVO患者中,Cirrus SD - OCT与Stratus OCT、Spectralis OCT与Stratus OCT以及Spectralis OCT与Cirrus SD - OCT的测量结果之间一致性较低。所有三种设备均显示出较高的组内重复性,DR患者中Stratus OCT的ICC为98%,Cirrus SD - OCT为97%,Spectralis OCT为100%。RVO患者中Stratus OCT的ICC为97%,Cirrus SD - OCT为79%,Spectralis OCT为91%。结论。不同设备间测量结果的一致性较低。然而,TD和SD - OCT设备的组内重复性均较高。