We evaluated the body fluid (BF) module on the new Sysmex XN-1000 for counting blood cells.
One hundred and eighty-seven BF samples [73 cerebrospinal fluid (CSF), 48 continuous ambulatory peritoneal dialysis (CAPD), 46 ascites, and 20 pleural fluid] were used for method comparison between the XN-1000 and manual microscopy (Fuchs-Rosenthal chamber and stained cytospin slides) for counting red blood cells (RBCs) and white blood cells (WBCs) (differential).
Good agreement was found for counting WBCs (y=1.06x+0.09, n=67, R2=0.96) and mononuclear cells (MNs) (y=1.04x-0.01, n=40, R2=0.93) in CSF. However, the XN-1000 systematically counted more polymorphonuclear cells (PMNs) (y=1.48x+0.18, n=40, R2=0.99) compared to manual microscopy. Excellent correlation for RBCs >1×109/L (y=0.99x+116.56, n=26, R2=0.99) in CSF was found. For other fluids (CAPD, ascites and pleural fluid) excellent agreement was found for counting WBCs (y=1.06x+0.26, n=109, R2=0.98), MNs (y=1.06x-0.41, n=93, R2=0.96), PMNs (y=1.06x+0.81, n=93, R2=0.98) and RBCs (y=1.04x+110.04, n=43, R2=0.98). By using BF XN-check, the lower limit of quantitation (LLoQ) for WBC was defined at 5×106/L. Linearity was excellent for both the WBCs (R2=0.99) and RBCs (R2=0.99) and carry-over never exceeded 0.05%.
The BF module on the XN-1000 is a suitable tool for fast and accurate quantification of WBC (differential) and RBC counts in CSF and other BFs in a diagnostic setting.
我们评估了新的 Sysmex XN-1000 上的体液(BF)模块,用于血细胞计数。
使用 187 个 BF 样本[73 份脑脊液(CSF)、48 份连续流动腹膜透析(CAPD)、46 份腹水和 20 份胸腔积液],比较 XN-1000 与手动显微镜(Fuchs-Rosenthal 室和染色细胞旋转涂片)在计数红细胞(RBC)和白细胞(WBC)(差异)方面的方法。
在 CSF 中,XN-1000 对 WBC(y=1.06x+0.09,n=67,R2=0.96)和单核细胞(MNs)(y=1.04x-0.01,n=40,R2=0.93)的计数具有良好的一致性。然而,与手动显微镜相比,XN-1000 系统地计数了更多的多形核细胞(PMNs)(y=1.48x+0.18,n=40,R2=0.99)。在 CSF 中,对于 RBC>1×109/L(y=0.99x+116.56,n=26,R2=0.99),XN-1000 与手动显微镜的相关性也非常好。在 CSF 中,XN-1000 与手动显微镜的 WBC(y=1.06x+0.26,n=109,R2=0.98)、MNs(y=1.06x-0.41,n=93,R2=0.96)、PMN(y=1.06x+0.81,n=93,R2=0.98)和 RBC(y=1.04x+110.04,n=43,R2=0.98)计数也具有良好的一致性。使用 BF XN-check,WBC 的定量下限(LLOQ)定义为 5×106/L。WBC(R2=0.99)和 RBC(R2=0.99)的线性都非常好,且携带污染从未超过 0.05%。
XN-1000 的 BF 模块是一种快速、准确地定量 CSF 和其他 BF 中 WBC(差异)和 RBC 计数的合适工具,适用于诊断环境。