In vivo accuracy of conventional and digital radiographic methods in confirming root canal working length determination by Root ZX.

机构信息

Department of Operative Dentistry, Endodontics and Dental Materials, Bauru School of Dentistry, University of São Paulo, Bauru, SP, Brazil.

出版信息

J Appl Oral Sci. 2012 Sep-Oct;20(5):522-5. doi: 10.1590/s1678-77572012000500005.

Abstract

OBJECTIVES

To compare, in vivo, the accuracy of conventional and digital radiographic methods in determining root canal working length.

MATERIAL AND METHODS

Twenty-five maxillary incisor or canine teeth from 22 patients were used in this study. Considering the preoperative radiographs as the baseline, a 25 K file was inserted into the root canal to the point where the Root ZX electronic apex locator indicated the APEX measurement in the screen. From this measurement, 1 mm was subtracted for positioning the file. The radiographic measurements were made using a digital sensor (Digora 1.51) or conventional type-E films, size 2, following the paralleling technique, to determine the distance of the file tip and the radiographic apex.

RESULTS

The Student "t" test indicated mean distances of 1.11 mm to conventional and 1.20 mm for the digital method and indicated a significant statistical difference (p<0.05).

CONCLUSIONS

The conventional radiographic method was found to be superior to the digital one in determining the working length of the root canal.

摘要

目的

比较传统和数字放射照相方法在确定根管工作长度方面的准确性。

材料与方法

本研究使用了 22 名患者的 25 颗上颌切牙或尖牙。考虑到术前射线照相作为基线,将 25 K 文件插入根管,直到 Root ZX 电子根尖定位仪在屏幕上指示 APEX 测量。从该测量值中减去 1 毫米以定位文件。使用数字传感器(Digora 1.51)或传统的 Type-E 胶片,尺寸 2,采用平行技术进行放射照相测量,以确定文件尖端和放射照相根尖之间的距离。

结果

Student“t”检验表明,传统方法的平均距离为 1.11 毫米,数字方法为 1.20 毫米,存在显著的统计学差异(p<0.05)。

结论

在确定根管工作长度方面,传统放射照相方法优于数字方法。