大视野镉锌碲硫(CdZnTe)SPECT/CT 相机进行铊心肌灌注成像的采集时间缩短:一项等效性研究。

Reduced acquisition time for thallium myocardial perfusion imaging with large field cadmium-zinc-telluride SPECT/CT cameras: An equivalence study.

机构信息

Service de Médecine Nucléaire, Hôpital Pierre Bérégovoy, Nevers, France.

Service de Médecine Nucléaire, CHU Saint-Etienne - Hôpital Nord, Saint Etienne, France.

出版信息

J Nucl Cardiol. 2022 Aug;29(4):1933-1941. doi: 10.1007/s12350-021-02611-z. Epub 2021 Apr 22.

Abstract

BACKGROUND

Cadmium-zinc-telluride (CZT) SPECT/CT cameras with large field of view offer a higher sensitivity than conventional Anger cameras. This prospective study aimed to determine the equivalence between a conventional protocol and a reduced acquisition time protocol for 201-Thallium myocardial perfusion imaging (MPI) using a whole-body CZT SPECT camera.

METHODS AND RESULTS

Stress MPI was obtained for 103 consecutive patients on a DISCOVERY-CZT camera. Images were anonymized and post-processed to simulate a 25% (D75 dataset) and 50% (D50 dataset) decrease in total recorded counts. Concerning the number of segments displaying a tracer uptake < 70% of maximum intensity per patient, equivalence was demonstrated for both count-reduced datasets with a good inter-observer agreement (between 0.90 and 0.88). When comparing the full-vs-D75 datasets and full-vs-D50 datasets, mean difference was 0.06 segment (CI95: [- 0.15;0.27], P < 0.001) and 0.518 segment (CI95: [0.28;0.76], P < 0.001) respectively. Inter-observer agreement was also moderate to good concerning the number of pathological segments (between 0.6 and 0.7) and excellent for functional parameters.

CONCLUSION

Whole-body CZT SPECT/CT cameras allow to reduce 201-Thallium MPI injected activity or acquisition time by 50% with an equivalence in the number of segments displaying a tracer uptake < 70% of maximum intensity and with a good inter-observer agreement.

摘要

背景

具有大视野的碲锌镉(CZT)SPECT/CT 相机比传统的闪烁相机具有更高的灵敏度。本前瞻性研究旨在确定使用全身 CZT SPECT 相机进行 201-铊心肌灌注成像(MPI)时,常规方案与减少采集时间方案之间的等效性。

方法和结果

在 DISCOVERY-CZT 相机上对 103 例连续患者进行了应激 MPI。对图像进行匿名处理并进行后处理,以模拟总记录计数减少 25%(D75 数据集)和 50%(D50 数据集)。就每位患者显示示踪剂摄取<70%最大强度的段数而言,两个计数减少数据集均表现出等效性,观察者间一致性良好(0.90 至 0.88)。比较全数据集与 D75 数据集以及全数据集与 D50 数据集,平均差异为 0.06 段(95%CI:[-0.15;0.27],P<0.001)和 0.518 段(95%CI:[0.28;0.76],P<0.001)。在观察到的病变段数(0.6 至 0.7)和功能参数方面,观察者间一致性也为中度至良好。

结论

全身 CZT SPECT/CT 相机可将 201-铊 MPI 注射活动或采集时间减少 50%,而不影响显示示踪剂摄取<70%最大强度的段数,且观察者间一致性良好。