Regional Institute of Ophthalmology, Minto Ophthalmic Hospital, BMCRI, Bengaluru, Karnataka, India.
Indian J Ophthalmol. 2021 Jul;69(7):1713-1717. doi: 10.4103/ijo.IJO_3364_20.
To compare central corneal thickness measurements obtained by Pentacam with those obtained by IOL Master 700, Cirrus Anterior segment optical coherence tomography and Tomey Specular microscopy in normal healthy eyes.
Two hundred and six eyes of healthy subjects were included in the study. Each subject was assessed by four different methods of measuring central corneal thickness using Pentacam, IOL Master 700, Cirrus AS-OCT and Tomey Specular microscopy by a single examiner.
The mean CCT [± standard deviation (SD)] for Pentacam, IOL Master 700, Cirrus AS-OCT and Tomey Specular microscopy were Pentacam (Oculus), AS-OCT (Cirrus), IOL Master 700 and Specular microscopy (Tomey) were 523.75 (±27.75), 525.29 (±28.81),517.13 (±28.43) and 512.82 (±27.60) μm, respectively. All the means were significantly different from one another (P < 0.000). The differences between pairs of mean central corneal thickness (CCT) for Pentacam and IOL Master, Pentacam and anterior segment- optical coherence tomography (AS-OCT), and Pentacam and Specular microscopy are statistically significant. Bland-Altman plots showed that pentacam and IOL Master 700 have the closest agreement, followed by AS-OCT. Specular microscopy was found to have the poorest agreement with Pentacam.
We found that CCT measurements of Pentacam did not correlate with measurements of IOL Master, or AS-OCT or Specular microscopy. In clinical practice, the devices analyzed should not be used interchangeably due to low agreement regarding CCT values.
比较 Pentacam 与 IOL Master 700、Cirrus 眼前段光学相干断层扫描和 Tomey 共焦显微镜在正常健康眼中测量的中央角膜厚度。
本研究纳入了 206 只健康眼。由同一位检查者使用 Pentacam、IOL Master 700、Cirrus AS-OCT 和 Tomey 共焦显微镜分别对每位患者的中央角膜厚度进行四次测量。
Pentacam、IOL Master 700、Cirrus AS-OCT 和 Tomey 共焦显微镜的平均 CCT(±标准差)分别为 Pentacam(Oculus)、AS-OCT(Cirrus)、IOL Master 700 和共焦显微镜(Tomey)523.75(±27.75)、525.29(±28.81)、517.13(±28.43)和 512.82(±27.60)μm,所有平均值彼此之间均有显著差异(P<0.000)。Pentacam 和 IOL Master、Pentacam 和眼前段光学相干断层扫描(AS-OCT)以及 Pentacam 和共焦显微镜的中央角膜厚度(CCT)平均值之间的差异均具有统计学意义。Bland-Altman 图显示,Pentacam 和 IOL Master 700 之间的一致性最好,其次是 AS-OCT。共焦显微镜与 Pentacam 的一致性最差。
我们发现 Pentacam 的 CCT 测量值与 IOL Master、AS-OCT 或共焦显微镜的测量值不相关。在临床实践中,由于 CCT 值的一致性较低,这些设备不应互换使用。