Department of Laboratory Medicine, University of Washington Medical Center, Seattle, Washington, USA.
J Clin Microbiol. 2013 Nov;51(11):3883-5. doi: 10.1128/JCM.02395-13. Epub 2013 Sep 18.
The results of the Focus Simplexa FluA/B & RSV Direct assay were compared to those of laboratory-developed reverse transcription PCR tests for 498 nasopharyngeal swabs. Concordance rates were 96.6% (476/493; κ = 0.91), 97.6% (481/493; κ = 0.47), and 99.2% (488/492; κ = 0.94) for influenza A, influenza B, and respiratory syncytial virus, respectively.
将 Focus Simplexa FluA/B & RSV Direct 检测的结果与实验室开发的 498 份鼻咽拭子逆转录 PCR 检测结果进行比较。一致性率分别为流感 A 为 96.6%(476/493;κ=0.91)、流感 B 为 97.6%(481/493;κ=0.47)和呼吸道合胞病毒为 99.2%(488/492;κ=0.94)。